Circular No. 44/2025/TT-BYT on providing regulations on the delegation of authority to implement some tasks and handle some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices

Category: Circular
  • Validity
  • Content
  • Diagram
  • Download

Validity Status

Circular No. 44/2025/TT-BYT on providing regulations on the delegation of authority to implement some tasks and handle some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices takes effect from 22/11/2025.
Effective from: 22/11/2025
Effective to: No expiration
Status: In effect

Content Details

MINISTRY OF HEALTH OF VIETNAM
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. 44/2025/TT-BYT

Hanoi, November 22, 2025

 

CIRCULAR

Providing regulations on the delegation of authority to implement some tasks and handle some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices

Pursuant to Law on Government Organization No. 63/2025/QH15;

Pursuant to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree NO. 04/2025/ND-CP;

Pursuant to Decree No. 217/2025/ND-CP dated August 05, 2025 of the Government on specialized inspection;

Pursuant to Decree No. 42/2025/ND-CP dated February 27, 02 of the Government on function, tasks, powers and organizational structures of the Ministry of Health of Vietnam;

At the request of the Director of the Infrastructure and Medical Device Administration (IMDA),

The Minister of Health hereby promulgates a Circular providing regulations on the delegation of authority to implement some tasks and handle some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices.

Article 1. Scope

This Circular provides for:

1. The delegation of implementation of some tasks and handling of some medical device-related administrative procedures under the jurisdiction of the Ministry of Health.

2. Templates of documents and reports on management of medical devices as specified in clause 12 Article 70 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

Article 2. Delegation of authority to implement some tasks and handle some medical device-related administrative procedures under the jurisdiction of the Ministry of Health to the IMDA

1. Issuance of new marketing authorization numbers of class C and class D medical devices as specified in clause 32, Article 32a of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

2. Inspection after issuance of marketing authorization numbers of class C and class D medical devices as specified in clause 3 and clause 4 clause 32 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

3. Suspension of circulation of  batches of medical devices, termination of suspension of circulation of batches of medical devices as specified in clause 3, Article 34 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

4. Recall of batches of medical devices as specified in clause 4 Article 34 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

5. Issuance of documents on the permission or prohibition of continued circulation of medical devices in accordance with the provisions of Article 37 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

6. Recall of numbers of marketing authorization of medical devices as specified in Article 39 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

7. Issuance of medical device import license as specified in Article 48 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

8. Issuance, renewal and revocation of certificates of free sale of medical devices as specified in Article 50 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

9. Issuance, supplementation and revocation of certificates of registration of inspection of medical devices as specified in Article 56 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

10. Decision on the handling of medical devices that fail to meet requirements for inspection as specified in Article 58 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP.

11. Specialized inspection for organizations and individuals engaged in the classification of medical devices; manufacturing, circulation, trading, exporting, importing and supply of medical device services; information and advertising of medical devices; management and use of medical devices at medical facilities in accordance with the provisions of Clause 1, Article 6 of Decree No. 217/2025/ND-CP.

Article 3. Templates of documents and reports on management of medical devices

Templates of documents and reports serving the management of medical devices as specified in clause 12 Article 70 of Decree No. 98/2021/ND-CP amended by Decree No. 07/2023/ND-CP, Decree No. 96/2023/ND-CP and Decree No. 04/2025/ND-CP are enclosed herewith:

1. Appendix I: Written declaration of, disclosure of, statement of, notification of or request for issuance of marketing authorization number, import license and certificate of free sale.

2. Appendix II: Personnel list.

3. Appendix III: Information on application for declaration, certificate, import license.

4. Appendix IV: Written commitment and request for continued circulation of medical devices.

5. Appendix V: Letter of authorization.

6. Appendix VI: Confirmation of eligibility for warranty.

7. Appendix VII: Technical document for medical devices.

8. Appendix VIII: Report on receipt, dispatch, inventory, use of medical devices, production materials, EQA reagents that contain narcotics/precursors.

Article 4. Entry into force

1. This Circular comes into force from November 22, 2025.

2. Circular No. 19/2021/TT-BYT dated November 16, 2021 of the Minister of Health shall be invalid from the effective date of this Circular.

Article 5. Transitional provision

1. Regarding applications that have been submitted to regulatory authorities before the effective date of this Circular and are made using the templates specified in Circular No. 19/2021/TT-BYT amended by Circular No. 10/2023/TT-BYT: If the such documents and reports are invalid or there are requests for amendments, organizations and individuals must use the templates enclosed herewith when updating or amending such documents and reports.

2. For Letters of authorization and Confirmations of eligibility for warranty of medical devices that are signed by the medical device owner for issuance using the templates in Circular No. 19/2021/TT-BYT amended by Circular No. 10/2023/TT-BYT before the effective date of this Circular and are still valid, they may be used continuously.

Article 6. Organization of implementation

Chief of the Ministry Office, Director of Department of Personnel and Organization, Director of Department of Legal Affairs, Director of IMDA, agencies, organizations and individuals are responsible for the implementation of this Circular.

Any difficulties arising during the implementation of this Circular should be reported to the Ministry of Health (via the IMDA) for timely consideration and handling./.

 

PP. MINISTER
DEPUTY MINISTER

Do Xuan Tuyen

 

APPENDIX I

TEMPLATES OF DECLARATION OF, DISCLOSURE OF, STATEMENT OF, NOTIFICATION OF AND REQUEST FOR ISSUANCE OF NUMBERS OF MARKETING AUTHORIZATION, IMPORT LICENSES AND CERTIFICATES OF FREE SALE
(enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health)

Template No. 01

Declaration of eligibility for production of medical devices

Template No. 02

Declaration of applied standards of class A medical devices

Template No. 03

Declaration of applied standards of class B medical devices

Template No. 04

Request for issuance of new marketing authorization numbers of class C and class D medical devices subject to national technical regulations

Template No. 05

Request for issuance of new marketing authorization numbers of medical devices of class C and class D medical devices that are measuring instruments whose samples require approval as prescribed by the law on measurement

Template No. 06

Request for issuance of new marketing authorization numbers of class C and class D medical devices eligible for quick issuance

Template No. 07

Request for issuance of new marketing authorization numbers of class C and class D medical devices eligible for emergency use authorization

Template No. 08

Request for issuance of new marketing authorization numbers of other class C and class D medical devices

Template No. 09

Declaration of eligibility for trading of class B, class C and class D medical devices

Template No. 10

Request for issuance of medical device import licence

Template No. 11

Request for issuance of certificate of free sale of class C and class D medical devices

Template No. 12

Notification of updated information in application for declaration of eligibility for production of medical devices

Template No. 13

Notification of updated information in application for declaration of applied standards of class A and class B medical devices

Template No. 14

Notification of updated information in marketing authorization application of class C and class D medical devices

Template No. 15

Notification of updated information in application for declaration of eligibility for trading of class B, class C and class D medical devices

Template No. 16

Disclosure of  planned medical device advertising contents and formats

Template No. 01

NAME OF MANUFACTURER
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

DISCLOSURE

OF ELIGIBILITY FOR PRODUCTION OF MEDICAL DEVICES

To: ………….2……………….

1. Name of manufacturing facility:

TIN: …………………………………………………………………..

Address of the head office: …….. 3 ………………………………………………………

Address of the facility: ………………….. 4……………………………………………

Tel: ………………………….. ………….. Fax: ………………………..

Email: ………………………………… Website (if any): …………..

2. Legal representative of the manufacturing facility:

Full name: ……………………………………………………………………

ID card number/passport number: ………..date of issue: ……………………. Place of issue: ……………………

Landline: ………………………. Mobile phone: ……………………

3. Medical devices produced by the manufacturing facility:

No

Name of medical device

Expected scale (products per year)

1.

2.

Declaration of eligibility for production of medical devices

Enclosures:

1.

ISO 13485 Certificate

□

2.

Documents regarding the system for monitoring and managing the processes of exporting, importing, inventory and usage of raw materials that are narcotics and precursors, as well as the processes of exporting, importing and inventory of medical devices containing narcotics and precursors, and the storage warehouse.

□

We hereby commit that:

1. The declared information is accurate and legal according to the regulations. We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. We ensure the fulfillment and maintenance of requirements declared during the process of operation.

3. Any changes related to the application for declaration of eligibility for production of medical devices will be updated as per regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 Department of Health of the province/central-affiliated city where the manufacturing facility is located

3 The address written on the business registration certificate

4 If it coincides with the registered business address, state “at the headquarters”

Template No. 02

FACILITY’S NAME
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

DECLARATION

OF APPLIED STANDARDS OF CLASS A MEDICAL DEVICES

To: ………….2……………….

1. Name of the declaring facility: ……………………………………………………………………

TIN or number of Licence for representative office establishment: ……………..

Address: ……3……… …………………………………………………………………….

Landline: .. ……………………………….. Fax:  .. ……………………………..

Email: …………………………………….……… ………………………………………..

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  …………………….

Landline: ………………………. Mobile phone: ………………………

3. Class A medical device:

Common medical devices4: □ or in vitro diagnostic medical devices: □

Name of medical device5: ………………………………………………………………………….

Trade name6 (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  ……………..

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: …………………………………………………………………..

Applied standards: ……………………………………………………………………..

4. For medical device that contains narcotic or precursor:

Name of narcotic/ precursor: …………………. Scientific name: ……………

CAS Registry Number (CAS RN): ……………………………………………………………………….

Concentration and content of the narcotic or precursor: ……………………………………..

Total content of narcotics or precursors in a smallest packing unit:

………………………………………………………………………………………………………….

5. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ……………………………………………………………………….

6. Information about warranty facility (if any):

Name of the facility: ……………………………………………………………………………………

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ……………………..

7. Number of declaration of eligibility for domestic production of medical devices:…

……………………………………………………………………………………………….

Declaration of applied standards of class A medical devices

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Documentation of brief technical description of the medical device in Vietnamese

□

5.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

6.

Technical documents in Vietnamese enclosed with documents on materials, product safety, manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

7.

Standard version declared to be applied by the medical device owner

□

8.

Certificate of conformity with standards

□

9.

For a medical device domestically manufactured: result of evaluating chemical, physical and microbiological parameters and other parameters granted by a qualified facility according to the provisions of law on conformity assessment or Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

10.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

11.

Template of label on the medical device

□

12.

Certificate of free sale for the imported medical device

□

We hereby declare that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the declaration.

3. Any changes related to the declaration will be updated in accordance with regulations.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 Department of Health of the province/central-affiliated city where the facility is located

3 The address written on the business registration certificate

4 A non-IVD medical device:

5 Name of the medical device in Vietnamese written according to the intended use of the medical device, except in cases where Vietnamese translation is not possible.

6  Trade name given by the medical device owner.

Template No. 03

FACILITY’S NAME
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

DECLARATION

OF APPLIED STANDARDS OF CLASS B MEDICAL DEVICES

To: ………….2……………….

1. Name of the declaring facility: ……………………………………………………………………

TIN or number of Licence for representative office establishment: ……………..

Address: ……3……… …………………………………………………………………….

Landline: .. ……………………………….. Fax:  .. ……………………………..

Email: …………………………………….……… ………………………………………..

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  …………………….

Landline: ………………………. Mobile phone: ………………………

3. Class B medical device:

Common medical devices4: □ or in vitro diagnostic medical devices: □

Name of medical device 5: ………………………………………………………………………….

Trade name6 (if any): ………………………………………………………………

Global Medical Device Nomenclature – GMDN code (if any):  ……………..

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: …………………………………………………………………..

Applied standards: ……………………………………………………………………..

4. For medical device that contains narcotic or precursor:

Name of narcotic/ precursor: ……………Scientific name: ………..……

CAS Registry Number (CAS RN): ……………………………………………………………………….

Concentration and content of the narcotic or precursor: ……………………………………..

Total content of narcotics and precursors in a smallest packing unit: …………………

………………………………………………………………………………………………..

5. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ……………………………………………………………………….

6. Information about warranty facility (if any):

Name of the facility: ……………………………………………………………………………………

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ……………………..

7. Number of declaration of eligibility for domestic production of medical device:…

……………………………………………………………………………………………….

Declaration of applied standards of class B medical devices

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Documentation of brief technical description of the medical device in Vietnamese

□

5.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

6.

Technical documents in Vietnamese enclosed with documents on materials, product safety, manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

7.

Standard version declared to be applied by the medical device owner

□

8.

Certificate of conformity with standards

□

9.

For a medical device domestically manufactured: result of evaluating chemical, physical and microbiological parameters and other parameters granted by a qualified facility according to the provisions of law on conformity assessment or Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

10.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

11.

Template of label on the medical device

□

12.

Certificate of free sale for the imported medical device

□

We hereby declare that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the declaration.

3. Any changes related to the declaration will be updated in accordance with regulations.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 Department of Health of the province/central-affiliated city where the facility is located

3 The address written on the business registration certificate

4 A non-IVD medical device:

5 Name of the medical device in Vietnamese written according to the intended use of the medical device, except in cases where Vietnamese translation is not possible.

6  Trade name given by the medical device owner.

Template No. 04

NAME OF REGISTERING FACILITY
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

WRITTEN REQUEST

for issuance of new marketing authorization numbers of class C and class D medical devices subject to national technical regulations

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the registering facility: ……………………………………………………………………

TIN or number of Licence for representative office establishment: ……………..

Address:………. 2……….. ………………………………………………………………………

Tel: ………………………………………Fax: …………………………………

Email: ……………………………………. ………………………………………………….

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ………… Date of issue: ………… Place of issue:  …………………

Landline: ………………………. Mobile phone: ………………………

3. Medical device requiring marketing authorization:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  ……………..

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Type of medical device: ………………………………………………………………………….

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: …………………………………………………………………..

4. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ……………………………………………………………………….

5. Information about warranty facility (if any):

Name of the facility: ……………………………………………………………………………………

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ……………………..

6. Number of declaration of eligibility for domestic production of medical device: ..

……………………………………………………………………………………………………..

7. Effective periods of documents contained in the application 3:

– ISO 13485 Certificate: …………

– Letter of authorization of the medical device owner. ……………………………..

– Certificate of free sale for the imported medical device: …………………………

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Certificate of free sale for the imported medical device

□

5.

Common Submission Dossier Template (CSDT)

□

6.

Certificate of conformity with regulations

□

7.

Documentation of brief technical description of the medical device in Vietnamese

□

8.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

9.

Technical documents in Vietnamese enclosed with documents on materials, product safety, product quality control and manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

10.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

11.

Template of label on the medical device

□

12.

Certificate of medical device clinical trial results issued by the competent authority in cases where clinical trials are required by law

□

We hereby delares that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the application.

3. Any changes related to the application will be updated in accordance with regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 In case of multiple documents: Specify expiry date of the document whose effective period is shortest.   In case of a document whose effective period is not specified, write “indefinite”.

Template No. 05

NAME OF REGISTERING FACILITY
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

WRITTEN REQUEST

for issuance of new marketing authorization numbers of medical devices of class C and class D medical devices that are measuring instruments whose samples require approval as prescribed by the law on measurement

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the registering facility: ……………………………………………………………………

TIN or number of Licence for representative office establishment: ……………..

Address:…………2………………………………………………………………..

Tel: ……………………………… ………Fax: …………………………………

Email: ……………………………………. ………………………………………………….

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: …… Place of issue:  …………………………….

Landline: ………………….. Mobile phone: ………………………

3. Medical device requiring marketing authorization:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Type of medical device ………………………………………………………………………….

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

4. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

5. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ………………………

6. Number of declaration of eligibility for domestic production of medical device: ..

7. Effective periods of documents contained in the application 3:

– ISO 13485 Certificate: …………

– Letter of authorization of the medical device owner: ……………………………..

– Certificate of free sale for the imported medical device: …………………………

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Certificate of free sale for the imported medical device

□

5.

Common Submission Dossier Template (CSDT)

□

6.

Decision on sample approval

□

7.

Documentation of brief technical description of the medical device in Vietnamese

□

8.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

9.

Technical documents in Vietnamese enclosed with documents on materials, product safety, product quality control and manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

10.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

11.

Template of label on the medical device

□

12.

Certificate of medical device clinical trial results issued by the competent authority in cases where clinical trials are required by law

□

We hereby delares that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the application.

3. Any changes related to the application will be updated in accordance with regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 In case of multiple documents: Specify expiry date of the document whose effective period is shortest.   In case of a document whose effective period is not specified, write “indefinite”.

Template No. 06

NAME OF REGISTERING FACILITY
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

WRITTEN REQUEST

for issuance of new marketing authorization numbers of class C and class D medical devices eligible for quick issuance

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the registering facility: … ………………………………………………………………..

TIN or number of Licence for representative office establishment: ……………..

Address:……2………….. …………………………………………………

Tel: ……………………………… ………Fax: …………………………………

Email: ……………………………………. …………………………………………

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: …… Place of issue:  …………………………….

Landline: ………………….. Mobile phone: ………………………

3. Medical device requiring marketing authorization:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Type of medical device ………………………………………………………………………….

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

4. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

5. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ………………………

6. Information about the circulation of the medical device:

– Number of document: …………………………………………………………………………

– Name of the issuing organization: …………………………………………………………………………

– Date of issue: ………………………………………………………………………………….

– Expiry date: ………………………………………………………………………

7. Number of declaration of eligibility for domestic production of medical device: ..

………………………………………………………………………………………………………

8. Effective periods of documents contained in the application3:

– ISO 13485 Certificate: …………

– Letter of authorization of the medical device owner: ……………………………..

– Certificate of free sale for the imported medical device: …………………………

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Certificate of free sale for the imported medical device

□

5.

Common Submission Dossier Template (CSDT)

□

6.

Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

7.

Documentation of brief technical description of the medical device in Vietnamese

□

8.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

9.

Technical documents in Vietnamese enclosed with documents on materials, product safety, product quality control and manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

10.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

11.

Template of label on the medical device

□

We hereby delares that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the application.

3. Any changes related to the application will be updated in accordance with regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 In case of multiple documents: Specify expiry date of the document whose effective period is shortest.   In case of a document whose effective period is not specified, write “indefinite”.

Template No. 07

NAME OF REGISTERING FACILITY
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

WRITTEN REQUEST

for issuance of new marketing authorization numbers of class C and class D medical devices eligible for emergency use authorization

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the registering facility: ……………………………………………………………………

TIN or number of Licence for representative office establishment: ……………..

Address:………. 2……….. ……………………………………………………………….

Tel: ……………………………… ………Fax: …………………………………

Email: ……………………………………. …… …………………………………………..

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: …… Place of issue:  …………………………….

Landline: ………………….. Mobile phone: ……………………..

3. Medical device requiring marketing authorization:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Type of medical device …………………………………………………………………………..

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

4. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

5. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ………………………

6. Information about circulation, emergency use authorization of medical devices (in case of technology transfer or processing where information about the transferred or processed products in circulation has to be provided):

– Number of document: …………………………………………………………………………

– Name of the issuing organization: …………………………………………………………………………

– Date of issue: ………………………………………………………………………………….

– Expiry date: ………………………………………………………………………

– Links to the documents about the circulation and permission for use of the medical device:  …..

7. Number of declaration of eligibility for domestic production of medical device:

……………………………………………………………………………………………………..

8. Effective periods of documents contained in the application3:

– ISO 13485 Certificate: …………

– Letter of authorization of the medical device owner: ……………………………..

– Certificate of free sale for the imported medical device: …………………………

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

CFS/MA or licence for emergency use authorization for the imported medical device.

□

5.

Document containing links to the applicant for issuance of marketing authorization numbers

□

6.

Common Submission Dossier Template (CSDT)

□

7.

Technology transfer contract

□

8.

Processing contract

□

9.

Certificate of product quality inspection or assessment

□

10.

Documentation of brief technical description of the medical device in Vietnamese

□

11.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

12.

Technical documents in Vietnamese enclosed with documents on materials, product safety, product quality control and manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

13.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

14.

Template of label on the medical device

□

15.

Certificate of medical device clinical trial results issued by the competent authority in cases where clinical trials are required by law

□

We hereby delares that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the application.

3. Any changes related to the application will be updated in accordance with regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 In case of multiple documents: Specify expiry date of the document whose effective period is shortest.   In case of a document whose effective period is not specified, write “indefinite”.

 

Template No. 08

NAME OF REGISTERING FACILITY
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

WRITTEN REQUEST

for issuance of new marketing authorization numbers of other class C and class D medical devices

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the registering facility: …. ……………………………………………………………….

TIN or number of Licence for representative office establishment: ……………..

Address:………. 2……….. ……………………………………………………………….

Tel: ……………………………… ………Fax: …………………………………

Email: ……………………………………. …… …………………………………………..

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: …… Place of issue:  …………………………….

Landline: ………………….. Mobile phone: ………………………

3. Medical device requiring marketing authorization:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Packing specification (if any): …………………………………………………………..

Type of medical device ………………………………………………………………………….

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

4. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

5. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ……………………………………………………………………………………….

Landline: ………………….. Mobile phone: ………………………

6. Number of declaration of eligibility for domestic production of medical device:

……………………………………………………………………………………………………………..

7. Effective periods of documents contained in the application3:

– ISO 13485 Certificate: …………

– Letter of authorization of the medical device owner: ……………………………..

– Certificate of free sale for the imported medical device: …………………………

Enclosures:

1.

ISO 13485 Certificate

□

2.

Letter of authorization of the medical device owner.

□

3.

Confirmation of eligibility for warranty

□

4.

Certificate of free sale for the imported medical device

□

5.

Common Submission Dossier Template (CSDT)

□

6.

The CSDT inspection record

□

7.

Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

8.

Certificate of analysis of composition and content of substances with antibacterial activity; assessment form for bio-efficacy of products for chemicals and preparations serving the only purpose of sterilization.

□

9.

Documentation of brief technical description of the medical device in Vietnamese

□

10.

Technical documentation of functions and specifications of the medical device promulgated by its owner

□

11.

Technical documents in Vietnamese enclosed with documents on materials, product safety, product quality control and manufacturing process and clinical and preclinical investigation reports including reports on the stability of IVD reagents, calibrators and controls.

□

12.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

13.

Template of label on the medical device

□

14.

Certificate of medical device clinical trial results issued by the competent authority in cases where clinical trials are required by law

□

We hereby declare that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the application.

3. Any changes related to the application will be updated in accordance with regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 In case of multiple documents: Specify expiry date of the document whose effective period is shortest.   In case of a document whose effective period is not specified, write “indefinite”.

 

Template No. 09

FACILITY’S NAME
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

DECLARATION

of eligibility for trading of class b, class c and class d medical devices

To: ………….2……………….

1. Name of the facility: …………………………………………………………………………………………

TIN: ………………………………………………………………………

Address:………3………………………………………

Transaction office (if any): …………………………………………………..

2. Legal representative of the facility:

Full name: ………………………………………………………………………..

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  ……………………….

Landline: …………………………. Mobile phone: ……………………

3. Technical staff of the business facility4:

Full name: ………………………………………………………………………..

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  ……………………….

Professional qualification: ……………………………………………………………

4. The list of medical devices/groups of medical devices traded by the facility:

…………………………………………………………………………………………………………….

…………………………………………………………………………………………………………….

…………………………………………………………………………………………………………….

…………………………………………………………………………………………………………….

Declaration of eligibility for trading of medical devices

Enclosures:

1.

Personnel list

□

2.

Records of storage warehouse for medical devices

□

3.

Records of means of transport of medical devices

□

4.

Documentary evidence of the storage warehouse and system for monitoring and managing the exporting,  importing and inventory process of medical devices containing narcotics or precursors

□

We hereby declare that:

1. The declared information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The quality of the medical device will be ensured and the medical device will be circulated in accordance with the declaration.

3. Any changes related to the application for declaration of eligibility for trading will be updated as per regulations.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 Department of Health of the province/central-affiliated city where the facility is located

3 The address written on the business registration certificate

4 Making detailed declaration based on the existing persons

 

Template No. 10

Name of importing organization or individual
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………….

……1………, ……[date]…., 20……

 

Written request for issuance of medical device import licence

To: Infrastructure and Medical Device Administration (IMDA)

Name of importing organization or individual: …………………………………………………….

TIN or ID card number/Passport number: …………………………………………..

Legal representative: ………………………………………………………………

Phone number: ………………………………………………………………………..

We hereby apply for the license to import the following medical devices:

No.

Name of medical device

Model/ code of product

Manufacturer, country of manufacture

Owner, country of ownership

Distributor, country of distribution
(if any)

Quantity

Unit

1. Purpose of import: …………………………………………………………………

2. User: ………………………………………………………………………..

3. We hereby declare:

– We will be responsible for ensuring the quality, model and quantity of imported medical devices

– The use of imported medical devices for their intended purposes will be ensured.

We will be responsible before law for breach of the above-mentioned commitments.

 

Importing organization or individual
Signature (full name, title) with confirmation by seal or digital signature

__________________________

1 Geographic name

Template No. 11

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

……1………(date) ………………., 20…

WRITTEN REQUEST

for issuance of certificate of free sale of class C and class D medical devices

To: Infrastructure and Medical Device Administration (IMDA)

1. The facility applying for a CFS:

1.1. Name of the facility: ………………………………………………………………..

1.2. The headquarter address: ………………… 2……………………………………….

1.3. Phone number: ……………………. Fax: …………………………………………

2. The medical device owner:

2.1. The owner’s name: ………………………………………………………….

2.2. Address: …………………………………………………………………..

3. Manufacturing facility:

3.1. Manufacturing facility 1:

– Name: ………………………………………………………

– Address: ………….. ………………………………………………

3.2. Manufacturing facility 2:

– Name: ……………………………………………………….

– Address: ………….. ………………………………………………

To meet the requirements of the importing country, our company requests the IMDA of the Ministry to Health to issue a CFS for the following medical devices:

No

Name of medical device

Model

Product code (if any)

Number of marketing authorization

Manufacturing facility3

1

2

3

Number of printed copies of the CFS requested for issuance:

We hereby declare:

1. We included adequate documents in the application and ensure their accuracy, legitimacy and conformity with regulations. We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The fulfillment and maintenance of requirements declared during the process of manufacture will be ensured.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name

2 The address written on the business registration certificate

3 Numbering manufacturing facility according to section 3

 

Template No. 12

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

…… 1………(date) ………………., 20…

NOTIFICATION

OF UPDATED INFORMATION IN APPLICATION FOR DECLARATION OF ELIGIBILITY FOR PRODUCTION OF MEDICAL DEVICES

Last updated on: ………….

To: Department of Health …..

1. Number of declaration: ……………………. Date: ….………………………

2. Updated information:

No

Out-of-date information

Up-to-date information

3. Related  documents: ………………

We hereby declare:

We will strictly comply with regulations of Vietnamese law;

We will be responsible before the law for the accuracy of the updates.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name.

Template No. 13

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

…… 1………(date) ………………., 20…

NOTIFICATION

OF UPDATED INFORMATION IN APPLICATION FOR DECLARATION OF APPLIED STANDARDS OF CLASS A AND CLASS B MEDICAL DEVICES

Last updated on: ………….

To: Department of Health …..

1. Number of the applied standard declaration: ………………. issued on: ………

2. Updated information:

No

Out-of-date information

Up-to-date information

3. Related  documents: ………………

We hereby declare:

We will strictly comply with regulations of Vietnamese law;

We will be responsible before the law for the accuracy of the updates.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature 

__________________________

1 Geographic name.

Template No. 14

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

…… 1………(date) ………………., 20…

NOTIFICATION

OF UPDATED INFORMATION IN APPLICATION FOR MARKETING AUTHORIZATION OF CLASS C AND CLASS D MEDICAL DEVICES

Last updated on: ………….

To: Infrastructure and Medical Device Administration (IMDA)

1. Number of marketing authorization: ………. ….Date of issue: ………….……

2. Updated information:

No

Out-of date information

Up-to-date information

3. Related  documents: ………………

We hereby declare:

We will strictly comply with regulations of Vietnamese law;

We will be responsible before the law for the accuracy of the updates.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name.

Template No. 15

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

…… 1………(date) ………………., 20…

NOTIFICATION

of updated information in application for declaration of eligibility for trading of class B, class C and class D medical devices

Last updated on: ………….

To: Department of Health …..

1. Number of declaration: ………….…. Date: ………………

2. Updated information:

No

Out-of-date information

Up-to-date information

3. Related  documents: ………………

We hereby declare:

We will strictly comply with regulations of Vietnamese law;

We will be responsible before the law for the accuracy of the updates.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature 

__________________________

1 Geographic name.

Template No. 16

FACILITY’S NAME
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………

……1………(date) ………………., 20…

 

DECLARATION

OF PLANNED MEDICAL DEVICE ADVERTISING CONTENTS AND FORMATS

To: Infrastructure and Medical Device Administration (IMDA)

1. Name of the facility2: ………………………………………………………………..

TIN: ………………………………………………………………………………….

Address: ……………………………………………………………………………………….

Tel: ………………………….. ………….. Fax: …………………………..

Email: ………………………………… …………. Website (if any): ………..

2. Legal representative of the facility:

Full name: ……………………………………………………………………………………

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  …………………….

Landline: ………………….. Mobile phone: ………………………

3. Information on the medical device planned for advertisement:

– Name of the medical device: …………………………………………………………….

– Model: ……………………………………………………………………………

– Product code (if any): ……………………………………………………………

– Manufacturer: ……………………………………………………………………….

– Country of manufacture: ………………………………………………………………………

– Number of marketing authorization: ………………………………………………………………………….

– Features and functions: ……………………………………………………………..

– Name and address of holder of marketing authorization number: ………..

– Health warnings for users, storage conditions (if any):

…………………………………………………………………………………………………………

We hereby disclose the planned medical device advertising contents and formats as follows:

Planned advertising contents:

……………………………………………………………………………….

……………………………………………………………………………….

Planned advertising formats:

……………………………………………………………………………….

……………………………………………………………………………….

We hereby declare:

1. The disclosed information is accurate and legal according to the regulations.  We will wholly be responsible for any falsification and will face penalties according to regulations of law.

2. The advertising content complies with the planned advertising content that has been publicly disclosed and the application for disclosure of applied standards for class A or class B medical devices or the application for marketing authorization of class C or class D medical devices.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature
 

__________________________

1 Geographic name

2 The holder of marketing authorization number or the organization authorized in writing by the holder of marketing authorization number

 

APPENDIX II

PERSONNEL LIST TEMPLATE
(Enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health of Vietnam)

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

…….1….., ……[date]….., 20…

Personnel list

Name of the facility: ……………………………………………………………………………………………………………

Address: …………………………………………………………………………………………………………………

No.

Full name

Position

Professional qualification

Process of working in the medical device-related field

Training process

Work unit

Work period

Designated position

Main assigned tasks

Name of training institution

Training major

Qualifications

Training type

Training duration

1

2

3

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or digital signature 

__________________________

1 Geographic name.

 

APPENDIX III

TEMPLATE OF INFORMATION ABOUT APPLICATION FOR DECLARATION, CERTIFICATE, IMPORT LICENSE
(enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health)

Template No. 01

Information about application for declaration of eligibility for production of medical devices

Template No. 02

Information about application for declaration of applied standards of class a medical devices

Template No. 03

Information about application for declaration of applied standards of class B medical devices

Template No. 04

Marketing authorization of class c and class d medical devices

Template No. 05

Information about application for declaration of eligibility for trading of class b, class c and class d medical devices

Template No. 06

Medical device import license

Template No. 07

Certificate of free sale

Template No. 08

Information about application for declaration of concentration and content in materials for production of medical devices and EQA reagents that contain narcotics/precursors

Template No. 01

…1…
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

 

INFORMATION

about application for declaration of eligibility for production of medical devices

Number of declaration: …………….

Date of declaration: ………….

1. Name of manufacturing facility: ……………………………………………………………………

2. Address: ……………………………………………………………………………………

3. Tel: ………………………. Fax: ………………………………………………

4. Document number of the facility:…………………………………issued on: …………………….. …………

5. Name of medical device declared to be manufactured:

…………………. ……………………………………………………………………………….

6. Enclosures:

1.

Declaration of eligibility for production

□

2.

ISO 13485 Certificate

□

3.

Documents regarding the system for monitoring and managing the processes of exporting, importing, inventory and usage of raw materials that are narcotics and precursors, as well as the processes of exporting, importing and inventory of medical devices containing narcotics and precursors, and the storage warehouse.

□

__________________________

1 Department of Health of the province/central-affiliated city where the manufacturing facility is located

Template No. 02

…1...
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

 

INFORMATION

About Application for declaration of applied standards of class A medical devices

Number of declaration: …………….

Date of declaration: ………….

1. Name of the declaring facility: ……………………………………………………………………

2. Address: ……………………………………………………………………………………

3. Document number of the facility:…………………………………issued on:………………….. ……………………….

4. Class A medical device:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Specifications of packing: ………………………………………………………………………

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

Applied standards: ……………………………………………………………………..

5. For medical device that contains narcotic or precursor (if any):

Name of narcotic/ precursor: …………………. Scientific name: ……………

CAS Registry Number (CAS RN): ……………………………………………………………………….

Concentration and content of the narcotic or precursor: ……………………………………..

Total content of narcotics and precursors in a smallest packing unit:

………………………………………………………………………………………………………….

6. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

7. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ………………………………………………………………………………………

Landline: ………………….. Mobile phone: ……………………..

8. Number of declaration of eligibility for domestic production of medical device:

.. ………………………………………………..…………………………………………..

9. Enclosures:

1.

Written request for declaration of applied standards of class A medical devices

□

2.

ISO 13485 Certificate

□

3.

Letter of authorization of the medical device owner.

□

4.

Confirmation of eligibility for warranty

□

5.

Standard version declared to be applied by the medical device owner

□

6.

Certificate of conformity with standards

□

7.

For a medical device manufactured in Vietnam: adding result of evaluating chemical, physical and microbiological parameters and other parameters granted by a qualified facility according to the provisions of law on conformity assessment, or Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

8.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

9.

Template of label on the medical device

□

10.

Certificate of free sale for the imported medical device

□

_________________________

1 Department of Health of the province/central-affiliated city where the facility is located

 

Template No. 03

…1…
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

 

INFORMATION

ABOUT APPLICATION FOR DECLARATION OF APPLIED STANDARDS OF CLASS B MEDICAL DEVICES

Number of declaration: …………….

Date of declaration: ………….

1. Name of the declaring facility: ……………………………………………………………………

2. Address: ……………………………………………………………………………………

3. Document number of the facility:…………………………………issued on:………………….. ……………………….

4. Class B medical device:

Name of the medical device: …………………………………………………………………………..

Trade name (if any): ……………………………………………………………….

Global Medical Device Nomenclature – GMDN code (if any):  …………….

Model: ………………………………………………………………………………….

Product code (if any): ………………………………………………………………….

Specifications of packing: ………………………………………………………………………

Uses: ………………………………………………………………………..

Name of manufacturing facility: ……………………………………………………………………….

Address of the manufacturing facility: ………………………………………………………………….

Applied standards: ……………………………………………………………………..

5. For medical device that contains narcotic or precursor (if any):

Name of narcotic/ precursor: …………………. Scientific name: ……………

CAS Registry Number (CAS RN): ……………………………………………………………………….

Concentration and content: …………………………………………………………………….

Total content of narcotics and precursors in a smallest packing unit: ………………………..

6. Information about the owner of the medical device:

The owner’s name: ……………………………………………………………………………

Address of the owner of medical device: ………………………………………………………………………

7. Information about warranty facility (if any):

Name of the facility: …………………………………………………………………………………..

Address: ………………………………………………………………………………………

Landline: ………………….. Mobile phone: ……………………..

8. Number of declaration of eligibility for domestic production of medical device:

.. ………………………………………….

9. Enclosures:

1.

Written request for declaration of applied standards of class B medical devices

□

2.

ISO 13485 Certificate

□

3.

Letter of authorization of the medical device owner.

□

4.

Confirmation of eligibility for warranty

□

5.

Standard version declared to be applied by the medical device owner

□

6.

Certificate of conformity with standards

□

7.

For a medical device manufactured in Vietnam: adding result of evaluating chemical, physical and microbiological parameters and other parameters granted by a qualified facility according to the provisions of law on conformity assessment, or Certificate of quality assessment issued by a competent authority of Vietnam for in vitro diagnostic medical devices

□

8.

User manual of the medical device in Vietnamese; for an imported medical device, an original in English, issued by the owner of the medical device, will be enclosed with the user manual in Vietnamese

□

9.

Template of label on the medical device

□

10.

Certificate of free sale for the imported medical device

□

__________________________

1 Department of Health of the province/central-affiliated city where the facility is located

 

Template No. 04

Ministry of Health
Infrastructure and Medical Device Administration
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. ………………..

Hanoi,….[date]..…., 20…

 

Marketing authorization of class C and class D medical devices

Pursuant to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Decree No. 07/2023/ND-CP dated March 03, 2023 of the Government on amendments to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Decree No. 96/2023/ND-CP dated December 30, 2023 of the Government elaborating on certain Articles of the Law on Medical Examination and Treatment;

Pursuant to Circular No. /2025/TT-BYT dated ……….., 2025 of the Minister of Health promulgating a Circular providing regulations on the delegation of implementation of some tasks and handling of some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices.

The Infrastructure and Medical Device Administration of the Ministry of Health hereby grants a marketing authorization to the following brand-new medical device:

1. Name of the medical device:

2. Trade name (if any):

3. Global Medical Device Nomenclature – GMDN code (if any):

4. Type of medical device:

5. Name and address of owner of medical device:

6. Name and address of holder of marketing authorization number:

7. Uses; name and address of warranty facility: See the attached appendix.

8. Detail on medical device (model; product code; packing specification; name of manufacturing facility; address of manufacturing facility and country of manufacture):  See attached appendix.

This Certificate is granted in accordance with Decision No………

 

AUTHORITY AND POSITION OF SIGNER
(signature of competent person, seal of issuer)

Full name

ENCLOSED APPENDIX

1. Uses:

2. Name, address of warranty facility:

3. Details on medical device:

No

Name of medical device

Model

Product code (if any)

Packing specification (if any)

Name of manufacturing facility

Address of manufacturing facility

Country of manufacture

Page……/……

Template No. 05

…1…
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

 

INFORMATION

about application for declaration of eligibility for trading of class B, class C and class D medical devices

Number of declaration: …………….

Date of declaration: ………….

1. Name of business facility: ………………………………………………………………………….

………………………………………………………………………………………………………….

2. Address: ………………………………………………………………………………………….

3. Document number of the facility:…………………………………issued on:………………….. …………………………….

4. Enclosures:

1.

Declaration of eligibility for trading

□

2.

Personnel list

□

3.

Records of storage warehouse of medical devices

□

4.

Records of means of transport of medical devices

□

5.

Documentary evidence of the storage warehouse and system for monitoring and managing the exporting, importing and inventory process of medical devices containing narcotics and precursors

□

__________________________

1 Department of Health of the province/central-affiliated city where the applicant is located

 

Template No. 06

Medical device import license

MINISTRY OF HEALTH
INFRASTRUCTURE AND MEDICAL DEVICE ADMINISTRATION
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. ………………..

Hanoi,….[date]..…., 20…

 

To: ……………………………………………

Pursuant to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Decree No. 07/2023/ND-CP dated March 03, 2023 of the Government on amendments to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Decree No. 96/2023/ND-CP dated December 30, 2023 of the Government elaborating on certain Articles of the Law on Medical Examination and Treatment;

Pursuant to Circular No. /2025/TT-BYT dated ……….., 2025 of the Minister of Health promulgating a Circular providing regulations on the delegation of implementation of some tasks and handling of some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices.

Considering the request in Official Dispatch No. …………. dated…………… on ………….. The Infrastructure and Medical Device Administration of the Ministry of Health hereby comes to an agreement with the request of (the applicant) on import of consignment of brand-new medical devices. To be specific:

No.

Name of medical device

Model/ code of product

Manufacturer, country of manufacture

Owner, country of ownership

Distributor, country of distribution
(if any)

Quantity

Unit

…..1 ……… shall be responsible for site inspection of goods quality before receipt, cooperate with Customs authority in solving import-related issues in accordance with regulations, be responsible for quality and safety of such goods, and use for intended purposes.

 

AUTHORITY AND POSITION OF SIGNER
(Signature of competent person, seal of issuer)

Full name

__________________________

1 Name of importing organization or individual.

Template No. 07

VIET NAM INFRASTRUCTURE AND MEDICAL DEVICE ADMINISTRATION
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

Hanoi, date … month … year …

CERTIFICATE OF FREE SALE

1. Certificate No: ………….. /CFS/HTTB

2. Product Owner:

Address:

3. Information of medical devices: (Product name; Model; Product code; Market Authorization number in Vietnam and Manufacturing site): See attached annex.

This is to certify that the above product(s) complies with the Medical Device regulations of Vietnam and is (are) allowed to be sold in Vietnam.

This certificate is issued according to Decision No: …….

 

Authority and position of the signer
(signature of the competent person, seal of the issuer (organization or agency))

Full name

Page …/….

Attached Annex

Certificate of Free sale No:
…………………………..CFS/HTTB

I. Manufacturing site(s):

1. Name of Manufacturer:

Address:

2. Name of Manufacturer:

Address:

II. Name of Product(s):

1. Product name:

Model:

Product code:

Market Authorization number in Vietnam:

Name of Manufacturer

2.

Product name:

Model:

Product code:

Market Authorization number in Vietnam:

Name of Manufacturer:

Page …/….

Template No. 08

…1…
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

INFORMATION

about application for declaration of concentration and content in materials for production of medical devices and EQA reagents that contain narcotics/precursors

Number of declaration: …………….

Date of declaration: ………….

1. Name of the declaring facility: ……………………………………………………………………………..

TIN: ……………………………………………………………………………………………

Address: ………………………………………………………………………………

Transaction office (if any): …………………………………………………………………

2. Legal representative of the facility:

Full name: ………………………………………………………………………..

ID card number/Passport number: ……. Date of issue: ……… Place of issue:  …………….

Landline: …………………………. Mobile phone: ……………………

3. Information on materials used for production of medical devices and eqa reagents that contain narcotics/precursors:

Product name: ………………………………………………………………………………….

Model: …………………………………………………………………………………………

Product code (if any): ………………………………………………………………………..

Packing specification (if any): …………………………………………………………………

Uses: ……………………………………………………………………………….

Applied standards: …………………………………………………………………………….

Name of manufacturing facility: ………………………………………………………………………………

Address of the manufacturing facility: …………………………………………………………………………

The owner’s name: …………………………………………………………………………………..

Address of the owner of medical device: ……………………………………………………………………………..

4. Information on narcotics/precursors contained in materials used for production of medical devices and EQA reagents:

Name of narcotic/ precursor: …………………………………………………………………..

Scientific name: …………………………………………………………………………..

CAS code ……………………………………………………………………………………………

Concentration and content: ……………………………………………………………………………

Total content of narcotics and precursors in a smallest packing unit: ……………………………….

5. Enclosures:

1.

Declaration of materials used for production of medical devices and EQA reagents that contain narcotics/precursors

□

2.

Certificate of quality (CQ)

□

3.

Technical document

□

__________________________

1 Department of Health of the province/central-affiliated city where the applicant is located

 

APPENDIX IV

TEMPLATE OF WRITTEN COMMITMENT AND APPLICATION FOR CONTINUED CIRCULATION OF CLASS C AND CLASS D MEDICAL DEVICES
(enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health)

Template No. 01

Written request for continued circulation of class C and class D medical devices and written commitments to take responsibility for warranty, preservation, provision of materials serving the use of medical devices

Template No. 02

Written consents for continued circulation of class C and class D medical devices

Template No. 01

Facility’s name
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. …………….

……1… …..,……(date)……..

 

Written request for continued circulation of class C and class D medical devices and written commitments to take responsibility for warranty, preservation, provision of materials serving the use of medical devices

To: Infrastructure and Medical Device Administration (IMDA)

Name of the facility: …………………………………………………………………………………..

TIN: ………………………………………………………………………………….

Address: …………………………………………………………………………….. ………….

Legal representative: ………………………………………………………………

Phone number: …………………………………………………………………………

We are ….2…. . We are distributing the following medical devices:

Name of the medical device: ……………………………………………………………………..

Number of marketing authorization: …………………………. Date of issue: ………………………………………..

Although ………….3…………. is unable to continue operations, we still have the capacity to ensure the quality of the above medical devices. Therefore, we are writing this letter to formally request the IMDA affiliated to the Ministry of Health to allow the continued circulation of the above medical devices and hereby declare that:

– We will be responsible for ensuring the quality, model and number of medical devices granted marketing authorization number.

– The medical devices will be circulated on the market for a maximum period of 24 months.

– We will be responsible for the warranty, maintenance and servicing of such medical devices.

– We will provide materials and replacement accessories during the 8-year period of use.

– We have adequate and qualified personnel in charge of effectiveness and safety of medical devices for users and environment, and ensure that the quality of the imported medical devices is not affected during transport and in our facilities. The medical device labels meet requirements.

– Medical devices are used for their intended purposes. We will facilitate inspections by competent authorities.

We will be responsible before law for breach of the above-mentioned commitments.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or digital signature
 

__________________________

1 Geographic name.

2 Name of distributing facility.

3 Name of the medical device owner or name of the holder of the marketing authorization number.

 

Template No. 02

Written consents for continued circulation of class C and class D medical devices

MINISTRY OF HEALTH
INFRASTRUCTURE AND MEDICAL DEVICE ADMINISTRATION
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. ………………..

Hanoi,….[date]..…., 20…

 

To: ……………………………………….

Pursuant to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Decree No. 07/2023/ND-CP dated March 03, 2023 of the Government on amendments to Decree No. 98/2021/ND-CP dated November 08, 2021 of the Government on management of medical devices;

Pursuant to Circular No. /2025/TT-BYT dated ……….., 2025 of the Minister of Health promulgating a Circular providing regulations on the delegation of implementation of some tasks and handling of some medical device-related administrative procedures under the jurisdiction of the Ministry of Health and regulations on templates of documents and reports on the management of medical devices.

Considering the content of the proposal in Official Dispatch No. …….. dated ……. from ……………. on the request to continue the circulation of medical devices and the commitment to assume responsibility for warranty, maintenance and supply of materials for the use of medical devices, the IMDA of the Ministry of Health agrees with the proposal of the organization or individual regarding the circulation of medical devices. To be specific:

Name of the medical device: ………………………………………………………..

Number of marketing authorization: ……………………. Date of issue: …………………………….

The product is allowed to be circulated until the end of: ……………………………………………………

The company………. assumes responsibility for:

– ensuring the quality, model and quantity of medical devices granted marketing authorization number.

– providing warranty, maintenance and servicing of such medical devices.

– Providing materials and replacement accessories during the 8-year period of use.

– Ensuring that we have qualified personnel in charge of effectiveness and safety of medical devices for users and environment and the quality of the imported medical devices is not affected during transport and in our facilities. Meeting requirements for medical device labels.

– Ensuring that medical devices are used for their intended purposes. We will facilitate inspections by competent authorities.

 

POSITION OF SIGNER
Signature (full name, title)
Confirmed by seal or signature
 

 

APPENDIX V

LETTER OF AUTHORIZATION TEMPLATE
(Enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health of Vietnam)

Headline (name and address) of the medical device owner

…[date]……, 20 ………………..

LETTER OF AUTHORIZATION

To: ………………………………………………….

We, (name and address of the Product Owner), as the owner of the medical devices listed hereunder, hereby authorize (name and address of the organization authorized to announce applicable standards in its name or register for circulation) to place/ register the following medical devices to the market of Vietnam:

“……….(List of the medical device)………”

We hereby commit to provide and support any inquiry related to the information and quality of the medical devices, guarantee all warranty, maintenance and service conditions and supply replacement materials and accessories for the medical devices.

This Letter of Authorization is valid until: ….. date (dd/mm/yy)

 

Legal representative of the owner
Signature (full name, title)
Confirmed by seal or signature
 

 

APPENDIX VI

CONFIRMATION OF ELIGIBILITY FOR WARRANTY TEMPLATE
(Enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health of Vietnam)

Headline (name and address) of the medical device owner: ……….

…[date]……, 20 ………………..

Confirmation of eligibility for warranty

First name: ………………………………………………………………………………. …………..

Address:…………………………………………………………………………..

As the owner of the medical device, we confirm that the below-named facilities are eligible for warranty on the medical device of …..1……………:

Name of medical device

Name of warranty facility

TIN

Address

Landline

Mobile phone

……

Facility 1

Facility 2

……

Facility 1

Facility 2

Facility 3

……

……

 

Legal representative of the owner
Signature (full name, title)
Confirmed by seal or signature 

__________________________

1 Full name of the owner of the medical device

 

APPENDIX VII

TEMPLATES OF TECHNICAL DOCUMENTS FOR MEDICAL DEVICES
(enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health)

Template No. 01 Technical documents for medical devices that are IVD reagents, calibrators, controls
Template No. 02  Brief technical description of chemicals and preparations serving solely the purpose of disinfecting medical devices
Template No. 03 Brief technical description of domestically manufactured medical devices which are neither IVD reagents, calibrators, controls nor chemicals and preparations serving solely the purpose of disinfecting medical devices
Template No. 04 Brief technical description of imported medical devices which are neither IVD reagents, calibrators, controls nor chemicals and preparations serving solely the purpose of disinfecting medical devices
Template No. 05 Common Submission Dossier Template (CSDT)
Template No. 06 Brief technical description of medical devices serving the issuance of import licenses

Template No. 01

Technical documents for medical devices that are IVD reagents, calibrators, controls

Name and address of facility registering for circulation of medical device

………(date) ……………….., 20….

No

Heading

Description

I

General description of medical device

1.1

Overview description Description and introduction of the medical device, its purposes and any complementary products (if applicable).

1.2

The history of product launches State the name of the first country to be granted a license and the year of issuance.

1.3

Uses State the intended use or indications as specified on the label or in the user manual.

1.4

List of countries in which licenses have been granted List the countries in which licenses have been granted and the year of issuance.

1.5

The status of submitted licensing applications pending approval in various countries List the countries in which applications have been submitted but have not yet been granted approval.

1.6

Important information regarding the safety and effectiveness of product Provide reports on adverse reactions that have occurred and the corrective actions that have been taken since the product was placed on the market (within the past five years).

II

Description of the medical device

2.1

Description of the medical device Describe the operating principle, features, and technical specifications of the medical device.

2.2

User manual Brief instructions for use of the medical device as prescribed in the Instruction sheet or Information sheet of the medical device

2.3

Contraindications Provide information on cases for which the medical device must not be used for reasons of patient safety, for example, due to the patient’s medical history, physiological characteristics, etc., in accordance with the contents stated on the label of the medical device

2.4

Warnings and precautions Provide warnings and precautions when using the medical device, including preventive measures to protect patients from risks caused by the use of the medical device. They can be warnings about adverse effects or misuse and precautions

2.5

Adverse effects Provide adverse effects related to the use of the medical device that is recorded through clinical trials and after-sales service which have been previously performed for the medical device

2.6

Alternative method (if any) Present alternative methods to achieve the intended purpose.

2.7

Technical specifications The performance features and technical specifications include: detection limits, accuracy, sensitivity, specificity, reliability and other factors; additional specifications encompassing chemical, physical and biological properties, sterility, stability (shelf life), storage, transportation and packaging.

III

Manufacture/production of medical device

3.1

Manufacturer Identify the manufacturers involved in the production process and the quality management system applied

3.2

Assessment standards and methods for class C and D IVD medical devices Provide quality control information during the process of production and dispatch of products. Provide standards of finished products and assessment methods. Provide the certificate of analysis of finished products of at least one batch from the manufacturer.

3.3

Stability Summarize the investigation results and conclusions regarding the stability of the product under storage conditions over real-time periods and accelerated aging investigations/studies (if any), during transportation and during use after opening (if any).

IV

Investigation report

4.1

Preclinical and clinical investigation Summarize preclinical and clinical investigation results of the product

4.2

References for preclinical and clinical investigation List references (if any)

V

Assessment of medical device quality in Vietnam Quality assessment information provided by the competent authorities of Vietnam

We certify that the above-mentioned information is true. We will be responsible before law for the above-mentioned information.

 

Legal representative of the facility
(Signature (full name, title))
Confirmed by seal or signature 

Template No. 02

Brief technical description of chemicals and preparations serving solely the purpose of disinfecting

MEDICAL DEVICES

Name and address of facility registering for circulation of medical device

………(date) ……………….., 20….

No

Heading

Content

1

Description of medical device

1.1

Description of the medical device Required information:

a) Composition of the preparation:

– Main composition (antimicrobial substances): Content; Manufacturer; UN No. (if any).

– Other auxiliary components of the preparation along with their content.

b) Physicochemical properties of the preparation: Specific gravity (relative density) for liquids; Flammability, flash point; Corrosiveness; Storage stability; Acidity, alkalinity or pH; Compatibility with other preparations.

c) The biological efficacy of the preparation:

– The mechanism of action of the preparation for sterilization/disinfection/cleaning…

– Sterilization/disinfection/cleaning capabilities…, types of bacteria.

– Dosage.

– Method of use.

– The exposure time of the chemicals to the test microorganisms.

– The interval between uses (for preparations with residual effects)

– Diluent environment, if any (water, oil, etc.)

d) Form of preparation.

e) Preservation conditions.

f) Expiry date.

1.2

Purposes/Indications State the intended use or indications of the product as specified on the label or in the user manual.

1.3

User manual Summarize instructions on how to use, store and observe safety precautions when using the preparation… according to the Instruction sheet or the Information sheet of the medical device.

1.4

Contraindications Information regarding cases where the use of medical devices is not recommended for safety reasons for users or the environment, etc.; in accordance with the instructions on the product label and as approved in the country of manufacture (if applicable).

1.5

Warnings and precautions Warnings and precautionary statements to be considered when using:

a) The product’s biocompatibility assessments:

– Assessments of the toxicity of the product such as: acute oral toxicity (LD50); acute dermal toxicity (LD50); acute inhalation toxicity (LC50); eye irritation potential; skin irritation potential; allergenic potential.

– Information on ecological toxicity, degradability, and persistence.

b) Information related to acute toxicity; semichronic toxicity; chronic toxicity; genotoxicity, carcinogenicity; reproductive and developmental toxicity; environmental metabolism potential; and ecological toxicity of antimicrobial substances (not required for chemicals or preparations whose sole purpose is the disinfection of medical devices categorized as class B).

c) Medical information, poisoning symptoms, antidotes (if available).

d) Other information such as the methods for disposing of expired or unused preparations; methods for disposing of preparation packaging.

1.6

Adverse effects Information regarding adverse effects related to the use of the product, such as potential negative impacts that may occur in individuals when using the product…

2

Information of the product circulated in countries (if any) Information on the countries granting product marketing authorization, the date of issue; and the approved indications of the product in each country.

3

Information related to adverse reactions, corrective actions (if any) a) Provide information on the number of reports on adverse reactions related to the use of medical devices.

b) Corrective actions for ensuring the safety in the market.

4

Information on product recalls (if any) Information regarding product recalls since the products were released to the market.

The recall/after-sales adjustment measures that have been implemented in accordance with the requirements of the regulatory authorities of the respective countries.

5

Manufacturer Identify the manufacturers involved in the production process and the quality management system applied

6

Medical device quality assessments Results of the testing of composition and content of antimicrobial active substances; trial of bio-efficacy of products

We certify that the above-mentioned information is true. We will be responsible before law for the above-mentioned information.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature 

 

Template No. 03

Brief technical description of domestically manufactured medical devices which are neither IVD reagents, calibrators, controls nor chemicals and preparations serving solely the purpose of disinfecting medical devices

Name and address of facility registering for circulation of medical device

….[date]…., 20…

No

Heading

Content

I

General description of medical device

1.1

Overview description Overview description of medical device, including: Introduction of the medical device, its purposes and indications for use, and any new features if available (for example: use of nanotechnology, artificial intelligence, etc.).

1.2

The history of product launches List of countries in which the product was sold on the market, attached to the first year (if any) in which the product was sold on the market.

1.3

Uses Purposes, indications for use of the medical device as stated on the label or in the device’s instruction manual.

1.4

Information of the grant of marketing authorization in countries Information about marketing authorization of the products in EU countries, Japan, Canada, Australia (TGA), USA (FDA), UK, Switzerland, including the licensing status (approved, pending approval, rejected, unregistered/non-registration, etc.), purposes, indications, initial authorization date.

1.5

Important information regarding the safety and effectiveness of product Brief information on adverse events that have been occurred, corrective actions for ensuring the safety in the market that have been taken since the products were introduced to the market or over the latest 5 years.

If a medical device contains any of the following composition, that composition must be clearly specified:

– Human or animal cells or tissues or their derivatives intended for use in a non-viable form, e.g. artificial heart valves from pig or cat intestines, etc.;

– Cells, tissues or derivatives of microbial or recombinant origin, e.g. hyaluronic acid-based dermal fillers obtained by bacterial fermentation, etc.

– Containing radioactive or ionizing ingredients (e.g. X-ray); or non-ionizing (E.g. laser, ultrasound, etc.).

II

Description of the medical device

2.1

Description of the medical device Detailed description of characteristics of the medical device for explaining principles of action of the medical device, explaining basic science concepts that make up the fundamentals of the medical device. Description of components and accessories used to help the device operate as well as package. Full description of each functional component, material or raw material of the medical device, accompanied by a visual of the medical device in the form of a diagram, photograph or drawing, as appropriate.

2.2

User manual All necessary information provided from the product owner including procedures, methods, frequency, timing, quantities and preparations that must be followed for the safe use of the medical device.

2.3

Contraindications Information on cases for which the medical device must not be used for reasons of patient safety, for example, due to the patient’s medical history, physiological characteristics, etc., in accordance with the contents stated on the label of the medical device or user manual of the medical device.

General description of diseases or cases and patient groups for which medical devices should not be used for the purpose of diagnosis, treatment or mitigation of disease. Contraindications that are cases for which medical devices should not be used because the risks clearly outweigh the potential benefits.

2.4

Warnings and precautions Warnings about specific dangers that users need to know before using medical devices.

Warnings issued to users to take necessary precautions for the safe and effective use of the medical device.

Including actions to be taken to avoid effects to the patient/user, which may not be life-threatening or cause serious injury, but of which the user should be aware.  Precautions may also warn users about the adverse effects when using or incorrectly using the medical devices, which can be avoided if the precautions are taken.

2.5

Potential adverse effects These are unwanted and serious consequences (death, injury, or serious adverse events) that may occur to the patient/user, or side effects from the use of the medical device under normal conditions.

2.6

Alternative therapy Description of alternative procedures or actions for diagnosing, treating or mitigating the disease or condition for which the medical device is intended.

2.7

Information about materials – A list of materials of the medical device that come into direct contact with the body (E.g. Mucous membrane) and indirect contact with the body (e.g. circulation of extracellular fluid (ECF)), and their chemical, biological and physical characteristics shall be provided.

– Regarding medical devices that emit ionizing radiation, information about sources of radiation (e.g. radioactive isotopes) and materials used to protect users and patients from unexpected radiation must be provided.

– In cases where there are special requirements regarding the safety of a material such as impurities, level of residue and exposure to plasticizers such as Bis (2-ethylhexyl ethylhexyl) phthalate (DEHP), a certificate of conformity to relevant standards, a certificate of analysis, or a safety assessment report of such material must be added Depending on risks incurred when exposing to these materials, additional measures may be required, such as notifying users of the presence of these materials by labeling the on the product.

2.8

Other relevant specifications Characteristics of the functions and technical specifications regarding the operation of medical devices include: accuracy, sensitivity, specificity of medical measuring and diagnostic devices, reliability and other factors (if relevant); and other specifications including chemistry, physics, electricity, mechanics, biology, software, sterility, stability, storage, transportation and packaging.

III

Synopsis of design verification and validation

This section must summarize or refer to or include design verification data and design validation data, appropriate to the complexity and risk classification of the medical device.

Documents include:

● Certificates or declarations of conformity to the recognized standards applied by the product owner; and/or

● Summaries or test reports and evaluations based on other standards, methods and trials of the manufacturer, or other means to demonstrate compliance with the product standards.

IV

Clinical evidence

Providing a clinical evaluation report of the medical device. This evaluation may be carried out in the form of a systematic review of available references, based on clinical experience with that medical device or similar medical device, or may be carried out through clinical investigations. Clinical investigations are often necessary for high risk medical devices, or medical devices where there is little or no clinical experience.

A clinical evaluation report includes purposes and context of the clinical evaluation, input clinical data, data appraisal and analysis, conclusions about safety and effectiveness of the medical device.

The clinical evaluation report should contain sufficient information as a standalone document for a regulatory authority to review. The clinical evaluation report should provide an overview of:

– Technology used by such medical device, indications for use, statements about the safety and clinical effectiveness of that medical device, if any.

– Nature and scope of the clinical data to be evaluated.

– Clinical data and recognized standards proving the safety and effectiveness of the medical device.

V

Manufacture information

5.1

Manufacturer The names and addresses of the manufacturers participating in the processes of manufacture and sterilization (including third-parties).

5.2

Manufacturing process It is required to include information serving the general perception of the manufacturing process. No proprietary details are required. This information can be presented in the form of a manufacturing flow diagram that briefly describes the process of manufacture, in-process quality control, assembly, quality control and packaging of the final product.

If multiple manufacturers are involved in the manufacturing process to complete a product, it is necessary to clearly state which activities each manufacturer is involved in.

We certify that the above-mentioned information is true. We will be responsible before law for the above-mentioned information.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature 

 

Template No. 04

Brief technical description of imported medical devices which are neither IVD reagents, calibrators, controls nor chemicals and preparations serving solely the purpose of disinfecting medical devices

Name and address of facility registering for circulation of medical device

………(date) ……………….., 20….

No

Heading

Brief description

1

Description of medical device

1.1

Description of the medical device Provide brief prescription of operating principles and features and technical parameters of the medical device; specify that if the device uses new technologies, a description of such new technologies (e.g. nanotechnology) should be provided.

1.2

Catalog of parts and accessories Make a list of parts and accessories of the medical device

1.3

Purposes/Indications Specify purposes/indications of the medical device

1.4

User manual Provide brief instructions for use of the medical device as prescribed in the Instruction sheet or Information sheet of the medical device

1.5

Contraindications Provide information on the contraindication – means cases in which the medical device must not be used for reasons of patient safety, for example due to the patient’s medical history, physiological characteristics, etc.; that is in accordance with the approved content in the country in which marketing authorization is granted and is stated on the label of the medical device

1.6

Warnings and precautions Provide warnings and precautions when using the medical device, including preventive measures to protect patients from risks caused by the use of the medical device. They can be warnings about adverse effects or misuse and precautions

1.7

Adverse effects Provide adverse effects related to the use of the medical device that are recorded through clinical trials and after-sales service which have been previously performed for the medical device

2

Provide information on the product that has been circulated in countries (if any)

Provide information on the countries in which the certificate of marketing authorization of the product is granted, or the first country in which the certificate of marketing authorization of the medical device is granted

3

Provide indications that have been registered in other countries (if any)

Make a list of countries in which marketing authorization attached with the indications approved in these countries are granted; dates of issue of the marketing authorization

4

Notable information about the safety/operation of the medical device:

– Provide information on the number of reports on adverse effects related to the use of the medical device; measures for recall/adjustment of after-sales services implemented at the request of regulatory authorities of some countries;

– Provide the following information if the medical device contains such components:

● Human or animal cells or tissues or their derivatives intended for use in a non-viable form – e.g. artificial heart valves from pig or cat intestines, etc.;

● Cells, tissues and/or derivatives of microbial or recombinant origin – e.g. skin plumping products based on hyaluronic acid obtained by bacterial fermentation, etc.; Contain irritating, ionizing ingredients – e.g. X-ray; or non-ionizing – E.g. laser, ultrasound, etc.

5

Reports on clinical assessment of class C and D medical devices which are not in vitro diagnostic medical devices

5.1

Clinical assessment Reports on clinical assessment of the owners of the medical devices in Vietnamese or English

5.2

References of the clinical assessment List of references (if any)

We certify that the above-mentioned information is true. We will be responsible before law for the above-mentioned information.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature
 

 

Template No. 05

Common Submission Dossier Template (CSDT)

Name and address of facility registering for circulation of medical device

We certify that the below-mentioned information is true. We will be responsible before law for the below-mentioned information.

1. General description of medical device

1.1. Brief description of medical device

Overview description of medical device, including: Introduction of the medical device, its purposes and indications for use and any new features if available (for example: use of nanotechnology, artificial intelligence, etc.).

1.2. Sales history of product on the market

Provide a list of countries in which the product was sold on the market, attached to the first year (if any) in which the product was sold on the market.

1.3. Purposes, indications for use

Specify purposes, indications for use of the medical device as stated on the label or in the device’s instruction manual.

1.4. Information of the grant of marketing authorization in countries

Provide information about the status of marketing authorization of the product in EU countries, Japan, Canada, Australia (TGA), USA (FDA), UK, Switzerland, including the licensing status (approved, pending approval, rejected, unregistered/non-registration, etc.), purposes, indications, initial authorization date1.

This information may be presented in tabular form as follows:

No

Name of issuing country or issuing authority

Purposes/Indications

Marketing authorization status

Date of issue

1

2

3

Provide copies or links to online documents on the grant of marketing authorization in the above-mentioned countries (if any).

1.5. Important information regarding the safety and effectiveness of medical device

Provide brief information on adverse events that have been occurred, corrective actions for ensuring the safety in the market that have been taken since the products were introduced to the market or over the latest 5 years.

Provide brief information on adverse events that have been occurred, containing at least a description of the adverse events, number of adverse events or frequency of occurrence (total number of adverse events recorded per total number of products sold). This information may be presented in tabular form as follows:

No

Description of adverse event

Number or frequency of occurrence

1

2

3

Provide information on corrective actions for ensuring the safety in the market, containing at least date of occurrence, brief description of the event, and name of the country or territory that took corrective action. This information may be presented in tabular form as follows:

No

Date

Brief description of the event

Name of the country or territory

1

2

3

Provide a written confirmation from the products’ owner if there have been no adverse events, no corrective action since selling products or over the latest 5 years.

Indicate whether the medical device contains one of the following ingredients:

– Human or animal cells or tissues or their derivatives intended for use in a non-viable form – e.g. artificial heart valves from pig or cat intestines, etc.;

– Cells, tissues or derivatives of microbial or recombinant origin – e.g. hyaluronic acid-based dermal fillers obtained by bacterial fermentation, etc.

– Radioactive or ionizing ingredients (e.g. X-ray); or non-ionizing (E.g. laser, ultrasound, etc.).

2. Essential principles checklist (available in Vietnamese or English)

Provide a checklist of essential principles regarding the safety and effectiveness of medical devices issued by the medical device owner as prescribed in the Appendix enclosed to prove the product is suitable to the related essential principles.

If the medical device has been granted a marketing authorization in EU countries, an essential principles checklist may be made and submitted in accordance with EU regulations.

3.  Description of the medical device

3.1 Description and presentation of characteristics of the medical device

Provide a detailed description of characteristics of the medical device for explaining its operating principles, clarifying basic scientific concepts that make up the fundamentals of the medical device. Provide a description of components and accessories used to enable the operation and packaging of the device. Provide a full description of each functional component, material or raw material of the medical device, accompanied by a visual of the medical device in the form of a diagram, photograph or drawing, as appropriate.

3.2 Uses

State the intended use of the medical device matching the data provided by the product owner in the user manual as well as the operational capabilities of the medical device.

3.3 Indications

Provide a general description of the pathology or condition that the medical device diagnoses, treats, prevents or mitigates, including a description of the target patient group for which the medical device is designed to be used.

3.4 User manual

All necessary information provided from the product owner including procedures, methods, frequency, timing, quantities and preparations that must be followed for the safe use of the medical device.

3.5 Contraindications

Provide information regarding cases where the medical device is not to be used due to patient safety reasons, for example, due to the patient’s medical history, physiological characteristics, etc., in accordance with the contents stated on the label of the medical device or user manual of the medical device.

Provide a description of diseases or cases and patient groups for which medical devices are not to be used for the purpose of diagnosis, treatment or alleviation of disease. Contraindications are cases in which the use of medical devices is not permitted because the risks clearly outweigh the potential benefits.

3.6 Warning

Provide warnings about specific dangers that users need to know before using medical devices.

3.7 Precautions

Warn users to take necessary precautions for the safe and effective use of medical devices. This may include actions that need to be taken to avoid effects to the patients/users; such effects may not pose a life-threatening risk or cause serious injury but the users should be aware of them. Precautions may also warn users about the adverse effects when using or incorrectly using the medical devices, which can be avoided if the precautions are taken.

3.8 Potential adverse effects

These are unwanted and serious consequences (death, injury, or serious adverse events) that may occur to the patient/user, or side effects from the use of the medical device under normal conditions.

3.9 Alternative therapy

Describe alternative procedures or actions for diagnosing, treating or alleviating the disease or condition for which the medical device is intended.

3.10 Materials (which may be listed in English or Vietnamese)

Describe materials of the medical device and its characteristics for proving the conformity with relevant essential principles.

Regarding a non-IVD medical device:

A list of materials of the medical device that come into direct contact with the body (E.g. Mucous membrane) and indirect contact with the body (e.g. circulation of extracellular fluid (ECF)), and their chemical, biological and physical characteristics shall be provided.

Regarding medical devices that emit ionizing radiation, information about sources of radiation (e.g. radioactive isotopes) and materials used to protect users and patients from unexpected radiation must be provided.

In cases where there are special requirements regarding the safety of a material such as impurities, level of residue and exposure to plasticizers such as Bis (2- ethylhexyl) phthalate (DEHP), a certificate of conformity to relevant standards, a certificate of analysis, or a safety assessment report of such material should be added. Depending on risks incurred when exposing to these materials, additional measures may be required, such as notifying users of the presence of these materials by labeling the on the product.

For IVD reagents, calibrators, controls:

A list of all the materials used in the production of the product, including name of material and its role in the finished product shall be provided.

Information about the composition, biological characteristics, and origins of materials involved in the testing reactions: antigens, antibodies, enzymes, conjugates, PCR primers, probes, calibrators, controls, etc. shall be provided.

For medical devices containing substances requiring special control (narcotics or radiation-producing materials, either non-ionizing or ionizing), it is required to provide adequate information about the composition, content and role of these substances, for example, Buprenorphine in drug testing kits, Iodine-131 in radioimmunoassay kits, Phosphorus-32 (32P) -labeled DNA probes in Southern blot methods, etc.

For class C and class D IVD medical devices Information on the standards or certificate of analysis of the main materials involved in the reactions (excluding stabilizers) shall be provided. The information described for each active material (antigen, antibody, multiplex PCR, etc.) shall be provided.

3.11 Other relevant specifications

Characteristics of the functions and technical specifications regarding the operation of medical devices include: accuracy, sensitivity, specificity of medical measuring and diagnostic devices, reliability and other factors (if relevant); and other specifications including chemistry, physics, electricity, mechanics, biology, software, sterility, stability, storage, transportation and packaging for proving the conformity with relevant essential principles.

3.12 Other information

Other important characteristics that are not mentioned above for proving the conformity with relevant essential principles (for example: the biocompatibility of medical devices, etc.).

Notes:  in cases where the above-mentioned information is provided in the user manual, the user manual may be used for reference

4. Synopsis of design verification and validation (in Vietnamese or English)

This section must summarize, refer to or include design verification data and design validation data, appropriate to the complexity and risk classification of the medical device.

This document includes:

Certificates or declarations of conformity to the recognized standards applied by the product owner; and/or

Summaries or trial reports and evaluations based on other standards, methods and trials of the manufacturer, or other means to demonstrate compliance with the product standards.

E.g. If the essential principles checklist mentions that the manufacturer uses a recognized standard to demonstrate conformity with an essential principle, a Declaration of Conformity to that standard must be provided, or a certificate of conformity, and a summary of the trial data if that standard does not include performance requirements.

Summaries of data or trial reports and evaluation, depending on the complexity and classification of the risk level of medical devices, typically include:

The list and conclusions drawn from published reports related to the safety and efficacy of medical devices in accordance with the essential principles;

Technical tests;

Laboratory tests;

Biocompatibility tests;

Animal experiments;

Using under simulated conditions;

Software evaluation.

4.1 Preclinical investigations

Detailed information on all biocompatibility tests performed on materials used in the medical device shall be provided. All significantly different materials must be described. Description of tests, results and data analyses must be provided.

Physical preclinical testing data must be fully provided, as appropriate. A report must include objectives, methods, results and conclusions of the product’s owner for all physical investigations of the medical device and components. Physical testing shall be performed to forecast the adequate response of the medical device to physical pressure, adverse conditions and effects, long-term use and all known and probable causes of device failure.

Reports on preclinical investigations on animals used to demonstrate possible effects of the medical device on humans shall be provided. These investigations must be conducted in compliance with good laboratory practice (GLP). A report must include objectives, methods, results, analyses and conclusions of the product’s owner. Conclusions of each investigation must address the interactions of the medical device with animal fluids and tissues and the effectiveness of the medical device in experimental animal model(s). Rationales (and limitations) of choosing specific animal models should be discussed.

Evidences of electrical safety and electromagnetic compatibility shall be provided. For example, if the owner declares that the product meets the requirements of IEC 60601-1 and IEC 60601-1-2, a test summary report and/or certificate of conformity must be provided to demonstrate that the medical device meets these standards.

Regarding sterilized medical devices, sterilization validation reports must be provided. If a sterilant is toxic or produces a toxic residue (e.g., Ethylene Oxide), data and test methods must be provided to demonstrate that the amount of sterilant or residue is within acceptable limits.

Evidences of cybersecurity for network-connected, internet-connected or wireless medical devices shall be provided. Such as: analyses of risks and possibilities of cyber attacks, cyber security control methods, plans, processes or mechanisms for monitoring, promptly detecting and managing security-related threats during the life cycle of the medical device. Declarations from the manufacturer or the owner stating that the medical device, which is network-connected, internet-connected or wireless-connected, ensures cyber security and safety may be provided.

4.1.1 Software verification and validation investigations (if applicable)

The correctness of a software product is another important characteristic of the product that cannot be fully validated in a finished product. The product owner must provide evidence confirming the software design and development process. This information includes results of the verification, validation and testing of the software done internally and in user’s environments prior to official launch, for all different hardware configurations such as mentioned on the label or in the user manual, as well as representative data obtained from both trial environments.

4.1.2 Medical device containing biological materials

It is required to provide results of investigation demonstrating that measures for facing dangers from infectious agents have been fully implemented. This section will include results of removing viruses and known hazards. Concerns about screening of donors must be fully addressed and methods of obtaining human material should be fully described. Results of process validation are required to demonstrate that manufacturing processes have been applied to minimize biological risks.

4.2 Preclinical investigations for IVD reagents, calibrators, controls

4.2.1 Analytical performance

It is required to provide investigation reports on the analytical performance of the medical device, including: analytical sensitivity, analytical specificity, limit of detection (LOD), limit of quantitation (LOQ), linearity, range, accuracy, repeatability, interfering substances, robustness, etc.  Performance criteria will depend on each medical device.

Information on samples used for evaluation (negative controls, positive controls, baseline standards, calibrator panels, reagents, …).

Investigation reports should include objectives, methods, results, and conclusions of the investigation.  The results and conclusions should clearly demonstrate that the product has the characteristics suitable for its intended use.

4.2.2 Stability

It is required to provide stability investigation reports, including both real-time shelf-life stability and accelerated shelf-life stability (if applicable). In the event that real-time shelf-life stability investigations are not conducted but only accelerated shelf-life stability investigations are conducted, a comprehensive and reasonable explanation is required.

In-use stability investigation reports for products that are repeatedly used after opening shall be provided.

It is required to provide shipping stability investigation reports, in which the products are shipped under real or simulated conditions.

Stability investigation reports must include objectives, methods, results, and conclusions.

4.3 Clinical evidence

A clinical evaluation report of the medical device shall be provided. This evaluation may be taken in the form of a systematic review of available references, based on clinical experience with that medical device or similar medical device, or may be conducted through clinical investigations. Clinical investigations are often necessary for high risk medical devices, or medical devices where there is little or no clinical experience.

A clinical evaluation report should include purposes and context of the clinical evaluation, input clinical data, data appraisal and analysis, conclusions about safety and effectiveness of the medical device.

The clinical evaluation report should contain sufficient information as a standalone document for a regulatory authority to review. The clinical evaluation report should provide an overview of:

– Technology used by such medical device, indications for use, statements about the safety and clinical effectiveness of that medical device, if any.

– Nature and scope of the clinical data to be evaluated.

– Clinical data and recognized standards proving the safety and effectiveness of the medical device.

4.3.1 Data collected from available references

If the product’s owner uses existing investigations and references to prove the safety and effectiveness of the medical device, copies of these documents should be provided.

Clinical evidences of the effectiveness may include medical device-related investigations conducted in Vietnam or other countries. These evidences may be collected from related investigations published in international medical journals.  Documents on clinical evidences must include objectives, methods and results presented in a clear and meaningful context. Before reaching conclusions of results of clinical investigations, discussions in context with published documents are required.

4.3.2 Data collected from clinical experience

Clinical experience is clinical data obtained from clinical use of the medical device, not from clinical investigations. Clinical experience may be of the medical device or comparable devices.

Clinical experience may be obtained from the following data:

Post-market surveillance reports of the product’s owner, regulatory authorities, cohort investigations (may contain unpublished long-term investigation data on safety and effectiveness).

Data on adverse events occurred obtained from the product’s owner or from regulatory authorities.

Data obtained from patients using the medical device in the pre-market aid program.

Information on clinically relevant field corrective actions such as recalls, notifications, hazard warnings.

4.3.3 Clinical investigation data

A clinical investigation is a systematic study on or in human subjects, undertaken to assess the safety and effectiveness of a medical device.

The clinical investigation may be undertaken by the medical device’s owner or by a third party representing the owner. The clinical investigation should be designed, conducted and reported in conformity with ISO 14155 standards, Part 1 and 2, Clinical Investigations of Medical Devices for Human Subjects, or in conformity with an equivalent standard, and in compliance with regulations of the host country.

The clinical investigation must conform to the ethical standards of the Declaration of Helsinki (DoH).

For an IVD medical device: Clinical investigations are investigation conducted to establish or validate the clinical performance of IVD medical devices. Manufacturers must provide clinical evidence to support their clinical claims, including: diagnostic sensitivity (clinical sensitivity) and diagnostic specificity (clinical specificity).

5. Labels of medical device

This is the descriptive and informational product literature that accompanies the medical device, such as: any physician/user manuals, pack labeling. This section must summarize, refer to or include the following labeling data to the extent appropriate to the complexity and risk class of the medical device, which is generally considered as “labeling”:

● Labels on the medical device and its packaging;

● User manual;

● Any information and instructions given to the patient, including instructions for any procedure the patient is expected to perform (if applicable).

5.1 Label samples on the medical device and its packaging

Label samples including information on the product must be presented in print, text or images, provided or attached to one or more layers of packaging, including secondary packages and primary packages. If it is impossible to provide physical label samples (e.g. a large warning label attached to an X-ray machine), the information may be provided in an alternative form, such as an image or technical drawing.

Label samples must meet requirements in accordance with regulations of applicable laws. As for imported medical devices, the label samples must include primary and secondary labels in Vietnamese.

The label sample must include all products stated in the application for marketing authorization. In case the application is submitted according to the medical device family, a label sample as a representative sample with a note of differences between label samples of the medical device shall be submitted.

5.2 User manual in Vietnamese

The user manual includes instructions for the physician or end user to use the medical device safely and for its intended use. The user manual should include indications, contraindications, warnings or precautions, potential adverse effects, alternative therapies, and conditions of storage and use to maintain safety and effectiveness of the medical device.

For IVD medical devices, the user manuals include the following information (if applicable):

– Type and code (if any) of the IVD medical device.

– Uses:

• Detect or measure something;

• Its function (e.g. screening, monitoring, diagnosis or aid to diagnosis, prognosis, predicting);

• Use of the same medical device (testing machine) whether it is automated or not;

• Whether it is quantitative or semi-quantitative or qualitative or identificatory, etc.;

• The type of specimen(s) required (e.g. serum, plasma, whole blood, tissue biopsy, urine);

• Testing population.

– Instructions: the product is used for IVD.

– The intended user (lay person or professional, etc.);

– The principle of the assay.

– Description of reagents, calibrators, reference materials, controls and limitations associated with their use (e.g. intended for use with a specific testing machine only).

– A list of provided materials and a list of specially required materials that are not provided.

– For IVD medical devices used in conjunction with other medical devices and/or in combination with non-medical devices or instruments:

• Information for identifying these devices or instruments, including important performance characteristics.

• Information on known restrictions when combining with those devices or instruments.

– Special storage conditions (e.g. Temperature, light, humidity and the conditions under which medical devices are used, if applicable.

– The in-use stability may include storage conditions and the shelf life after the first opening; storage conditions and the stability of working solutions, if applicable. The product’s lifespan must be clearly stated in the user manual.

– If a medical device is provided in a sterile form, it is necessary to specify that the medical device is sterile, specify the sterilization method and the handling instructions in the event that the sterile packaging is damaged prior to use.

– The information that allows users to be informed about any warnings, precautions, necessary measures to be taken and usage restrictions for IVD medical devices. This information should include, if applicable:

• Warning, precautions and/or measures to be taken in the event that the IVD medical device is damaged or malfunction or degradation reflected through the change in the external form of the medical device may affect performance.

• Warnings, precautions and/or measures to be taken regarding exposure to foreseeable external influences or environmental conditions, such as external magnetic fields, electricity and electromagnetic forces, electric discharges, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, or temperature;

• Warnings, precautions and/or necessary measures to be taken regarding potential risks during the use of the medical device in specific diagnosis, evaluation and treatment (e.g., electromagnetic interference emitted by the device affecting other devices);

• Precautions related to materials incorporated into the medical device that contain carcinogens, mutagens or substances harmful to reproduction, endocrine disruptors, or substances that may cause sensitivity or allergic reactions in patients.

– Conditions for collecting, handling and preparing specimen.

– Instructions for preparing or handling medical devices before use, such as sterilization, assembly, calibration, etc., to ensure that the device is used in accordance with the owner’s intended purpose.

– The necessary information to determine when IVD medical device is properly installed and ready to operate safely and as intended by the owner, along with the following information, when applicable:

• Details about the nature and frequency of preventive maintenance and scheduled maintenance, including cleaning or disinfection;

• Consumables and how to replace them;

• Information regarding the calibration that needs to be performed to ensure that medical device operates correctly and safely throughout its life cycle;

• Measures to mitigate risks encountered by individuals involved in installation, calibration or repair, such as contaminated surfaces.

– Recommendations on the quality control process when necessary.

– Instructions on how to retrieve the values assigned to calibrators and controls, including information for identifying reference materials and the reference measurement procedures used.

– The process of conducting tests includes calculating and interpreting results and recommending additional confirmatory tests if appropriate.

– The analytical performance characteristics, such as sensitivity, specificity and precision (which is a combination of trueness and accuracy).

– Clinical performance characteristics, such as diagnostic sensitivity and diagnostic specificity when applicable.

– Reference range when applicable.

– Information about interfering substances or limits (e.g., visual evidence of hyperlipidemia or hemolysis, the specimen storage duration) may affect the execution of the analysis.

– Warnings or precautionary measures to be taken regarding the disposal of the device, its accessories and consumables used together with it, if any. This information should include the following contents, if applicable:

• Hazards related to infection or microorganisms (e.g., consumables contaminated with substances capable of transmitting infections of human origin);

• Environmental hazards (e.g., batteries or radiative materials that emit potentially hazardous radiation);

• Physical hazards (e.g., explosions).

– For IVD medical devices used by lay persons, it is necessary to advise users to seek the opinions of healthcare professionals before making any medical decisions.

– References, if applicable.

– The name and address of the owner must be presented in a recognizable format that allows for the identification of the owner’s information. In cases where the user manual contains information about the manufacturing facility and origin, this information must include complete details regarding the name and address of the manufacturing facility.

– Document identification information, such as version number or release date.

6. Risk analysis (written in either Vietnamese or English)

This section must summarize, refer to or include the results of the risk analysis. This risk analysis is based upon international standards or other recognized standards, and is appropriate to the complexity and risk classification of the medical device.

Results of risk analysis

A list of possible risks for the medical device shall be provided. Indirect risks from the medical device may originate from hazards related to the medical device itself, such as moving parts causing long-term injuries, or from hazards related to the user, such as ionizing radiation from an X-ray machine. Assessments of these risks in relation to the declared benefits of the medical device and methods used to reduce risks to an acceptable level must be presented. Individuals or organizations analyzing risks must be identified clearly. Techniques used to analyzing risks must be defined to ensure the appropriateness to the medical device and the risks involved.

7. Information on manufacture (in Vietnamese or English)

This section must summarize, refer to or include documents related to the manufacturing processes, including quality assurance measures, in accordance with the complexity and risk classification of medical devices.

7.1 Manufacturer’s information

The names and addresses of the manufacturers participating in the processes of manufacture and sterilization (including third-parties).

7.2 Manufacturing process

It is required to include information serving the general perception of the manufacturing process. No proprietary details are required. This information can be presented in the form of a manufacturing flow diagram that briefly describes the process of manufacture, in-process quality control, assembly, quality control and packaging of the final product.

If multiple manufacturers are involved in the manufacturing process to complete a product, it is necessary to clearly state which activities each manufacturer is involved in.

 

Legal representative of the applicant
Signature (full name, title)
Confirmed by seal or digital signature
 

__________________________

1For products granted a certificate of marketing authorization in EU countries, the latest grant date may be written if the initial authorization date is not stated.

APPENDIX

Essential principles checklist

(in English or Vietnamese)

Name of the owner of the medical device:

Name of the medical device:

Product code (if any):

Essential principles

Applicable to the medical device?

Applied standards

Documentary evidence

General requirements

1. Medical devices must be designed and manufactured so that when they are used under conditions and for the intended purposes, and with the knowledge and experience of users, they will not cause damage to health or safety of patients, users or others, provided that any risks associated with the use of these medical devices for their intended purposes are acceptable risks when considering benefits for patients and in accordance with requirements for health and safety protection at a high level.

2. Solutions of medical device owners for design and manufacture of medical devices must be in accordance with safety principles and take into account the level of scientific and technical development. During the process of selection of suitable solutions for design and manufacture of the medical devices, in order to minimize risks associated with the use of these medical devices, the owners must:

– Identify hazards and associated risks arising from the use of the medical devices and the foreseeable misuse,

– Eliminate or minimize risks reasonably through safe design and manufacturing,

– For risks that cannot be eliminated, fully implement protective measures, including warnings where necessary, and

– Inform users of residual risks.

3. Medical devices must achieve the effectiveness intended by the medical device owners and must be designed, manufactured and packaged in a manner suitable for one or more functions within the definition of a medical device.

4. The technical characteristics mentioned in Clauses 1, 2 and 3 should not cause adverse effects to such a degree that the health or safety of patients or users and other persons is compromised throughout the expected life of the medical devices; when the medical devices are subjected to the stresses which can occur during normal conditions of use and have been maintained and calibrated according to the instructions of medical device owners.

5. Medical devices must be designed, manufactured and packaged in such a way that their technical characteristics, when used for their intended purposes, will not be adversely affected during transport and storage if the transport and storage are carried out in accordance with the instructions and information provided by the medical device owners

6. Benefits to be gained must outweigh any undesirable effects for the effectiveness intended.

7. Medical devices require clinical evidence, appropriate for their use and risk classification, demonstrating that the devices comply with the respective essential principles. The clinical evaluation must be conducted.

Requirements for design and manufacturing

8. Chemical, physical and biological properties

8.1 Medical devices must be designed and manufactured in such a way that requirements for technical characteristics mentioned in Clauses 1 to 6 of “General requirements” are satisfied.  Particular attention should be paid to:

– The selection of materials to be used, in particular regarding toxicity, flammability,

– Physical and chemical properties of the materials to be used,

– The compatibility between the materials to be used and biological tissues, cells, body fluids, and specimens, taking into account the intended purposes of the medical devices,

– Reasons for selection of the material, such as abrasion, hardness and resistance of metal.

8.2 Medical devices must be designed, manufactured and packaged in such a way that reduces risks caused by pollutants and residues for patients and relevant persons during transport, storage and use of the medical devices. Particular attention should be paid to tissues exposed and to the duration and frequency of exposure during the transport, storage and use of the medical devices.

8.3 Medical devices must be designed and manufactured in such a way that they can be safely used with the materials, substances and gases to which they are exposed during normal conditions of use or in routine procedures; If medical devices are used for infusion, they must be designed and manufactured in such a way that they are compatible with the medicinal products concerned and the action of such drugs is maintained for their intended use.

8.4 If a medical device contains a substance that, if it is used separately, could be considered a medicinal product under the relevant legislation and has the effect of supporting the action of the medical device on the body, the safety, quality and effectiveness of the medical device as well as the safety, quality and effectiveness of the combined medicinal product must be verified. The “medicinal product” referred to herein includes stable derivatives from human blood or plasma.

8.5 Medical devices must be designed and manufactured in a way that minimizes risks caused by substances that may seep or leak from the medical device.

8.6 Medical devices must be designed and manufactured in a way that minimizes risks posed by inadvertent ingress or release of substances from the medical devices, taking into account the nature of the environment in which they are intended to be used.

9. Infection and microbial contamination

9.1 Medical devices and manufacturing processes must be designed in a way that reasonably and appropriately eliminates or minimizes risks of infection to any person. The design must:

– Allow for easy handling, and, where necessary:

– Reasonably and appropriately minimize any microbial leakage and/or microbial exposure during use;

– Minimize the transfer of contamination from patients, users or other persons to the medical devices or specimens and vice versa during use.

9.2 Where a medical device is composed of substances of biological origin, the risk of infection must be reasonably and appropriately minimized by selecting appropriate sources, sample donors and substances, and using validated inactivation, conservation, testing and control procedures, when appropriate. This requirement may not be applicable to certain in vitro diagnostic medical devices if viral activity and infectious agents are essential to the intended uses of the devices or if being removed or inactivated, the effectiveness of such in vitro diagnostic medical devices will be affected.

9.3 Medical devices made from non-living tissues, cells and substances of animal origins will be considered to have animal origin under relevant regulations, and thus be subject to veterinary control and supervision for the intended use of these tissues. The medical device owners should maintain information about the geographical origin of animals.  The processing, storage, inspection and handling of tissues, cells and substances of animal origin must be carried out in a manner that ensures optimal safety. In particular, the safety related to viruses and other infectious agents must be achieved by validated methods of removal or inactivation during the manufacturing process. This requirement may not be applicable to certain in vitro diagnostic medical devices if other viral activity and infectious agents are essential to the intended use of the products or if being removed or inactivated, the effectiveness of such in vitro diagnostic medical devices will be affected.

9.4 For products made from cells, tissues and derivatives of bacterial or recombinant origin that are determined to be medical devices, the selection of sample sources/donors, processing, preservation, testing and handling of cells, tissues and derivatives of such origin must be carried out in a manner that achieves optimal safety. In particular, the safety related to viruses and other infectious agents must be achieved by validated methods of removal or inactivation during the manufacturing process. This requirement may not be applicable to certain in vitro diagnostic medical devices if other viral activity and infectious agents are essential to the intended use of the products or if being removed or inactivated, the effectiveness of such in vitro diagnostic medical devices will be affected.

9.5 For products composed of non-living tissues, cells, and substances of human origin that are determined to be in vitro diagnostic medical devices, the selection of sample sources, sample donors, and/or substances of human origin, the processing, storage, testing and handling of tissues, cells and substances of such origin must be carried out in a manner that ensures optimal safety. In particular, the safety related to viruses and other infectious agents must be achieved by validated methods of removal or inactivation during the manufacturing process. This requirement may not be applicable to certain in vitro diagnostic medical devices if other viral activity and infectious agents are essential to the intended use of the products or if being removed or inactivated, the effectiveness of such in vitro diagnostic medical devices will be affected.

9.6 Medical devices labeled as having a specific microbial state must be designed, manufactured and packaged to ensure that they remain in that state when placed on the market and remain so during transport and storage as specified by the medical device owners.

9.7 Medical devices delivered in a sterile state must be designed, manufactured and packaged to ensure that they remain sterile when placed on the market and remain sterile under the transport and storage conditions specified by the medical device owners.

9.8 Medical devices labeled as sterile or as having a specific microbial state must be processed, manufactured and, if necessary, sterilized by means of appropriate and validated methods.

9.9 Medical devices intended to be sterilized must be manufactured in appropriate and controlled conditions (e.g. environment).

9.10 Packaging for non-sterile medical devices must be capable of keeping the medical devices at the prescribed level of cleanliness. If medical devices need to be sterilized before use, the packaging system should minimize the risk of microbial contamination with consideration taken of sterilization methods of the medical device owners. Such medical devices must be manufactured in appropriate and controlled conditions.

9.11 Packaging and/or labeling of medical devices must be distinguishable from identical or similar products on the market in both sterile and non-sterile conditions.

10. Manufacturing and environmental properties

10.1 If a medical device is intended to be used in combination with other medical devices or tools, the whole combination, including the connection system should be safe and should not impair the specified effectiveness of the medical devices or tools. Any known restrictions on use applying to such combinations should be indicated on the label and/or in the instructions for use.

10.2 Medical devices should be designed and manufactured in such a way as to remove or reasonably and appropriately reduce the:

– Risks of injury in connection with their physical features and, as appropriate, mass/pressure ratio, dimensional and design characteristics;

– Risks connected with reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, pressure, humidity, temperature or variations in pressure and acceleration;

– Risks associated with the use of the medical devices when they come into contact with materials, substances or gases, to which they can be exposed during normal conditions of use;

– Risks from accidental penetration of substances into the medical devices;

– Risks of incorrect identification of samples;

– Risks of interference with other medical devices normally used in investigation or treatment;

– Risks arising from failure to maintain or calibrate (as with implantable devices), from aging of the materials used or from the loss of accuracy of some measurement or control mechanism.

10.3 Medical devices must be designed and manufactured in such a way as to minimize the risks of fire or explosion during normal use or in case of malfunction. Particular attention should be paid to medical devices whose intended use includes exposure to or use in association with flammable or combustible substances.

10.4 Medical devices should be designed and manufactured in such a way as to facilitate their safe disposal of waste substances.

11. Medical devices with a diagnostic or measuring function

11.1 Medical devices with a measuring function should be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose. The limits of accuracy, correctness and stability should be indicated by the medical device owners.

11.2 Medical devices shall be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose, based on appropriate scientific and technical methods. In particular, the design must address sensitivity, specificity, trueness, repeatability, reproducibility, known confounding factor control and detection limits, if any.

11.3 Where the operation of medical devices depends on the use of calibrators and/or controls, traceability of values assigned to such calibrators and/or controls must be ensured through a quality management system.

11.4 Any measurement, monitoring or display scale must be designed in accordance with scientific design principles, taking into account purposes of the medical devices.

11.5 Whenever possible, values expressed numerically should be in commonly accepted, standardized units, and understood by users of the medical devices.

12. Protection against radiation

12.1 General

Medical devices should be designed, manufactured and packaged in such a way that exposure of patients, users and other persons to radiation is appropriately minimized in a manner that is compatible with the intended purpose, whilst not restricting the application of levels for diagnostic and therapeutic purposes.

12.2 Intentional radiation

12.2.1 Where medical devices are designed to emit hazardous or potentially hazardous radiation that is visible or invisible to users for specific medical benefits that outweighs its inherent risks, and users can control such radiation, these medical devices should be designed and manufactured to ensure the relevant variable parameters stay within a tolerance.

12.2.2 Where medical devices are intended to emit hazardous or potentially hazardous radiation, visible and/or invisible to users, they should be fitted with visual displays and/or audible warnings of such emissions.

12.3 Unintended radiation

Medical devices should be designed and manufactured in such a way that the exposure of patients, users and other persons to the emission of unintended, stray or scattered radiation is appropriately reduced as far as possible.

12.4 User manual

The operating instructions for medical devices emitting radiation should contain detailed information as to the nature of the emitted radiation, the means of protecting the patients and users, and ways of avoiding misuse and of eliminating the risks during installation.

12.5 Ionizing radiation.

12.5.1 Medical devices emitting ionizing radiation should be designed and manufactured to ensure that the shape, quantity and distribution of energy (or quality) of such emitted radiation can be changed and controlled where possible, taking into account the intended use.

12.5.2 Medical devices emitting ionizing radiation intended for diagnostic X-rays should be designed and manufactured in such a way as to achieve an image and/or output quality that are appropriate to the intended medical purpose whilst minimizing radiation exposure of patients and users.

12.5.3 Medical devices emitting ionizing radiation intended for therapeutic X-rays shall be designed and manufactured in such a manner as to properly monitor and control doses used, beam types, energy and, where possible, monitor and control the energy distribution of radiant beams.

13.  Requirements for medical devices connected to or equipped with an energy source.

13.1 Medical devices that incorporate programmable electronic systems, including software, should be designed to ensure the repeatability, reliability and operability of these systems according to their intended use. In the case of a single fault state in the system (SFC), appropriate means shall be used to eliminate or appropriately reduce the risks as for as possible.

13.2 For medical devices combined with medical software or being medical software by themselves, such software must be evaluated on an advanced scientific-technical basis, taking into account the principles of development life cycle, risk management, evaluation and verification.

13.3 Medical devices for which the patients’ safety depends on an internal power source must be equipped with a power status indicator.

13.4 Medical devices for which the patients’ safety depends on an external power supply must be equipped with an alarm system to signal when power goes out.

13.5 Medical devices intended to monitor one or more clinical parameters of patients must be equipped with an appropriate alarm system to alert users to situations that could lead to death or serious deterioration of the patients’ health status.

13.6 Medical devices shall be designed and manufactured to appropriately minimize as far as possible risks of generating large amounts of electromagnetism that could impair the operation of the medical devices themselves or other medical devices located nearby.

13.7 Medical devices must be designed and manufactured to ensure immunity to electromagnetic interference so that these devices can function as intended.

13.8 Protection against electrical risks

Medical devices must be designed and manufactured in such a way that patients or anyone else are protected in the best way against electrocution accidents when such devices are installed and maintained in accordance with the owners’ instructions, in a normal operating condition and in a single fault condition (SFC)

14. Protection against mechanical risks

14.1 Medical devices must be designed and manufactured to protect patients and users against mechanical risks connected with their use.

14.2 Medical devices shall be designed and manufactured to minimize risks arising from vibrations generated by the medical devices, taking account of technical progress and of the means available for limiting vibrations, particularly at source, unless the vibrations are part of the operation of the devices.

14.3 Medical devices shall be designed and manufactured to minimize risks arising from the noise emitted, taking account of technical progress and of the means available to reduce noise, particularly at source, unless the noise emitted is part of the operation of the devices.

14.4 Terminals and connectors to user-controlled electrical, gas, or hydraulic and pneumatic power supplies shall be designed and constructed to minimize any possible risks.

14.5 Accessible parts of medical devices (excluding the parts or areas intended to supply heat or reach given temperatures) and their surroundings should not attain dangerous temperatures under normal conditions of use.

15 Protection against risks posed to patients by medical devices supplying energy or substances

15.1 Medical devices for supplying patients with energy or substances should be designed and installed in such a way that the rate and/or amount to be delivered can be set and maintained accurately enough to ensure the safety of patients and users.

15.2 Medical devices should be fitted with means of preventing and/or indicating any inadequacies in the rate and/or amount of energy delivered or substances delivered which could pose a danger. Medical devices should be equipped with suitable means to prevent, as far as possible, the accidental release of energy or substances at dangerous levels.

15.3 The control and indicator display functions must be clearly specified on medical devices. Where a medical device bears instructions required for its operation or indicates operating or adjustment parameters by means of a visual system, such information should be understandable to users and, as appropriate, patients.

16. Active implantable medical devices

16.1 Active implantable medical devices must have clear information to determine:

– Categories of medical devices;

– The medical device owner;

– Year of manufacture of medical devices.

16.2 That information must be readable without surgery on transplant recipients.

17. Protection against risks for patients posed by medical devices for use or testing by lay users

17.1 These medical devices must be designed and manufactured in such a way as to ensure that they perform appropriately for their intended purpose, taking into account the skills and the means available to lay users and the influence resulting from variations that can be reasonably anticipated in the lay users’ technique and environment.   The information and instructions provided by the medical device owners should be easy for the lay users to understand and apply.

17.2 These medical devices must be designed and manufactured in such a way as to minimize risks of error during use, and if applicable, in the handling of testing samples and in the interpretation of testing results.

17.3 These medical devices should, where appropriate, include a means by which the lay users can verify that, at the time of use, the medical devices will perform as intended by the owners.

18. Information provided by medical device owners

Users must be provided with information necessary to identify medical devices, owners of medical devices, explanations for use of medical devices in a safe manner, etc.

19. Clinical investigation

Clinical investigations conducted on human subjects must be based on the Declaration of Helsinki.  This requirement covers all steps in a clinical investigation from initial consideration of the need and justification of the investigation to the publication of results.

Write “Yes”, if such essential principles are applicable to the medical devices;

Write “No”, if such essential principles are not applicable to the medical devices.

If one essential principle is not applicable to the medical devices, explanations are required.

E.g. if a medical device does not contain biological materials, in the column “applicable requirements”, write “No” and specify that the medical device does not contain biological materials.

 (List the reference standards or internal standards and procedures that the manufacturer applies to ensure the product meets the essential principles)

E.g.

– ISO 13485

– ISO 14971

– EN ISO 18113-1

– IEC 61326-1

– IEC 61326-2-6

– IEC 60601-1

(state the technical documentation that demonstrates compliance to the essential principles, E.g.  certificates, investigation reports, test results etc.)

Template No. 06

Brief technical description of medical devices serving the issuance of import licenses

Name of the importing facility (name, address)

………(date) ……………….., 20….

No

Heading

Brief description

1

Description of medical device

1.1

Prescription of the medical device Provide brief prescription of operating principles and features and technical parameters of the medical device; specify that if the device uses new technologies, a description of such new technologies (e.g. nanotechnology) should be provided.

1.2

Catalog of parts and accessories Make a list of parts and accessories of the medical device

1.3

Purposes/Indications Specify purposes/indications of the medical device

1.4

User manual Provide brief instructions for use of the medical device as prescribed in the Instruction sheet or Information sheet of the medical device

1.5

Contraindications Provide information on the contraindication – means cases in which the medical device must not be used for reasons of patient safety, for example due to the patient’s medical history, physiological characteristics, etc.; that is in accordance with the approved content in the country in which marketing authorization is granted and is stated on the label of the medical device

1.6

Warnings and precautions Provide warnings and precautions when using the medical device, including preventive measures to protect patients from risks caused by the use of the medical device. They can be warnings about adverse effects or misuse and precautions

1.7

Adverse effects Provide adverse effects related to the use of the medical device that are recorded through clinical tests and after-sales service which have been previously performed for the medical device

2

Information of the product circulated in countries (if any)

Provide information on the countries in which the marketing authorization of the product is granted, or the first country in which the marketing authorization of the medical device is granted

3

Indications that have been registered in other countries (if any)

Make a list of countries in which marketing authorization attached with the indications approved in these countries are granted; dates of issue of the marketing authorization

4

Notable information about the safety/operation of the medical device:

– Provide information on the number of reports on adverse effects related to the use of the medical device; measures for recall/adjustment of after-sales services implemented at the request of regulatory authorities of some countries;

– Provide the following information if the medical device contains such components:

● Human or animal cells or tissues or their derivatives intended for use in a non-viable form – e.g. artificial heart valves from pig or cat intestines, etc.;

● Cells, tissues and/or derivatives of microbial or recombinant origin – e.g. skin plumping products based on hyaluronic acid obtained by bacterial fermentation, etc.; Contain irritating, ionizing ingredients – e.g. X-ray; or non-ionizing – E.g. laser, ultrasound, etc.

We certify that the above information is true. We will be responsible before law for the above-mentioned information.

 

Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature
 

 

APPENDIX VIII

TEMPLATE OF REPORT ON RECEIPT, DISPATCH, INVENTORY, USE OF MEDICAL DEVICES, PRODUCTION MATERIALS, EQA REAGENTS THAT CONTAIN NARCOTICS/PRECURSORS
(enclosed with Circular No. 44/2025/TT-BYT dated November 22, 2025 of the Minister of Health)

FACILITY’S NAME
——-

SOCIALIST REPUBLIC OF VIETNAM
Independence – Freedom – Happiness
—————

No. ……

 

 

REPORT ON RECEIPT, DISPATCH, INVENTORY, USE IN …………
OF MEDICAL DEVICES AND EQA REAGENTS THAT CONTAIN NARCOTICS/PRECURSORS

To: .…………….………………………

No.

Name of medical device/ production material/ EQA reagent

Number of declaration

Narcotic/ precursor

Name of narcotic/ precursor

CAS code

Content in a smallest packing unit

Unit

Remaining quantity of the previous period

Received quantity

Used1

Dispatched quantity

Remaining quantity at the end of the current period

Loss 2

This report is enclosed with the detailed list of names and addresses of customers, and names and quantity of narcotics/ precursors dispatched in the year.

 

………….., day……..month……….year ……
Legal representative of the facility
Signature (full name, title)
Confirmed by seal or signature 

_________________________

1 Specify quantity of production materials.